X. INSTALLATION

 

1. Operation Environment: 

  1) Temperature

0℃

40℃

  Maximum Relative Humidity: 80% 

2

High Temperature: High Temperature will result in a mildewing, dewing and even ruinous 

instrument. 

3

Avoid  placing  the  instrument  in  a  dusty  environment.  When  ending  your  microscope 

operation, please cover it with the dust cap. 

4

Please lay the microscope in a plan and stable position. 

2

Input Power Checking: the supplied voltage should be accordant with the required voltage 

which marked in the bottom of the microscope, or the instrument will be damaged. 

3

Lamp: 

1

 

The lamp is already installed when the instrument left the factory. 

2

 

When using, or soon after it is turned off, the lamp and the cooling part is very hot, so 

you should keep yourself carefully from burn. 

3

 

Replacing the lamp

 

  When replacing the lamp, be sure the bulb is the appointed one at the first. Then turn the 

main switch on “O” (off), pull up power plug, waiting until the bulb, the lamp house and periphery 

are all cool down, so you can begin your replacing: lay down the microscope, pull up the bulb top 

control switch, turn out the top, then insert the bulb into the holder stably and safely, and return the 

top back to finish. 

Don’t  touch  the  Halogen bulb  with  your  hands. Please  do  the  replacement with  your 

glove or  the lamp  protected  cover.  If  you  leave fingerprints on  the surface carelessly, use a 

soft  cloth,  dipping  with  some pure  alcohol.  Then  clean  them  away.  If  you  didn't  do  so,  the 

stains will shorten the lamp's service life or cause it to burst.

 

4

Binocular Tube: 

Mount  the  binocular  tube  onto  the  polarizer  accessories  stayed  on  the  main  brocket  and 

screw down the setscrew. 

5

Eyepiece (ocular): 

Insert the eyepiece into the binocular tube. 

6

Objective: 

Rotating  the  stage  until  it  drop  to  the  lowest  limited  position,  then  screw  the  different 

magnification objectives in to the turret (nosepiece) and fix them in turn. 

7

Condenser: 

The  condenser  had  mounted  already  before  left  the  factory.  if  you  need  to  take  it  down, 

please loose the locking bolt. 

8

Polarizer: 

When the polarizer was in the optical path, its polarized direction is east-west at 0°. It had 

been adjusted and located on the collector before left the factory, so it didn't allow disassembled. 

When  the  polarizer  swung  out  of  the  optical  path,  the  system  is  stay  in  an  analyzed  (single 

polarized light) state. 

9

Color filters: 

        Place the filters on the iris diaphragm brocket under the collector when you use them.   

 

Содержание A15.1016

Страница 1: ...1 A15 1016 A15 1017 Polarizing Microscope Instruction Manual Please read this manual carefully before using the microscope and let it s along with the instrument...

Страница 2: ...r plate quartz wedge wedge wedge wedge Strain Free Objective Rotatable Stage Stage rotation clamping knob Polarizer Condenser Bertrand Lens Microscope Body Coarse Focus Knob Fine Focus Knob Power Swit...

Страница 3: ...32 dry 60 optional 0 85 0 17 2 969 0 185 dry 3 Crosshair Scale Ocular KIND MAGNIFICATI ON FOCUS MM LINEAR FIELD OF VIEW MM PLAN LENS 10 24 95 18 4 Sum Magnification OCULAR 10 10 10 10 OBJECTIV 4 10 40...

Страница 4: ...iece Quadruple Nosepiece Center Adjustable Optical Compensator slip first class red 1 4 slip quartz wedge blank slip Illumination Halogen Lamp 6V20W spare Lamp spare BGX 1 20 0 5A Fuse 2 2 Eyepiece Ey...

Страница 5: ...nto the holder stably and safely and return the top back to finish Don t touch the Halogen bulb with your hands Please do the replacement with your glove or the lamp protected cover If you leave finge...

Страница 6: ...l axis coincident with each other The nosepiece is center adjustable the objective optical axis and stage has been pre adjusted coincident at the factory and should not require readjustment If they ar...

Страница 7: ...ogonal polarized light state push in the bertrand lens and rise up the condenser then you can observe the convergent light performance of the specimen 7 Optical Compensator Constant optical compensato...

Страница 8: ...in pure alcohol or xylene Not use organic solvent to wipe the non optical elements when must clean please use the soft detergent note that the alcohol and the xylene are all burned easily do not let...

Отзывы: