BIOLINE
Manual for installation and use
OPTIKON 2000
9 CLEANING, STERILISATION AND MAINTENANCE
The medical personnel must take the responsibility to keep existing instruments and
equipment functioning properly. The following simple steps serve as practical
guidelines in establishing your care and maintenance program.
9.1 CLEANING
9.1.1 UNIT
1) Periodically clean the touch screen LCD with soft dusting cloth dampened with
distilled water. If necessary use only neutral detergent.
2) Do not use alcohol, solvents, or abrasives on the control console.
9.1.2 ACCESSORIES
Treat the accessories as indicated above
9.2 STERILISATION
9.2.1 UNIT
The console and footswitch do not need to be sterilised. For cleaning please refer to
the indications given in the relevant instructions for use.
9.3 MAINTENANCE
9.3.1 UNIT
1) The BIOMETER should be stored in a clean and dry environment at room
temperature.
NOTE:
To maintain system safety, the compliance to EN60601-1 leakage current limits should be tested at
least once a year. Please, inquire with the Bioengineering Department in your clinic or with your local
Optikon Distributor/Service Center.
Code 141002EN
9-1
26.10.05 Rev.0