Page 12
9. Use of the microscope
9.1 Microscope switch on
Put the main switch
①
to “I” (ON) to turn on the
microscope. (Fig. 19)
9.2 Light intensity adjustment
Rotate the light intensity adjustment knob
②
to
increase / decrease the illumination voltage. (Fig. 20)
9.3 Centering the bulb
1.
Remove one objective from the nosepiece and
put the empty hole in the optical path.
2.
Put a piece of paper on the stage, slightly unscrew
the knurled ring of the focus adjustment lever
③
on the lamp housing and rotate it counteclockwise
until the bulb filament is in focus on the paper.
(Fig. 21)
•
If the bulb filament is not in the center of the
light path, the resulting image will be not evenly
illuminated. In this case a re-centering is needed.
3.
Rotate the centering screws
④
on the right side
of the lamp housing to bring the bulb filament in
the center of the field of view.
4.
At the end fully rotate the focus adjustment lever
③
clockwise and screw the knurled ring to fix the
lever into its position.
5.
Reinstall the objective.
Fig. 19
Fig. 20
Fig. 22
Fig. 21
①
②
③
④