10. Use of the microscope in polarized light
•
The system allows observation in Orthoscopy (crossed Nicol) or in Conoscopy (crossed Nicol with
the use of the Bertrand lens).
• For optimal performance in polarized light microscopy, accurate optical adjustments are essential
before beginning the observation.
10.1 Centering the nosepiece
1.
Center the condenser as described in chapter
9.7.
2.
Loosen the stage rotation lock
①
and rotate the
stage until the graduated scale of the stage
②
and the vernier scale
③
are aligned in the “0”
position. (Fig. 14)
•
This operation serves to ensure a standard
reference position for centering the
nosepiece.
3.
Focus on a recognizable detail
④
in the field of
view by placing it at the center of the crosshair.
(Fig. 15)
4.
By turning the stage, the detail in focus will
describe a circle
⑤
. (Fig. 15)
5.
Move again the stage on the “0” position and
lock the locking screw
①
. Operating on the
nosepiece centering screws
⑥
move the detail in
diametrically opposite direction to the described
circle. The movement must be about half the
diameter of the circle described. (Fig. 16)
6.
Manually move the specimen and put it again
on the center of the crossahir. Loosen again the
stage rotation lock and rotate once more the
stage.
7.
If the centering has been carried out correctly, by
turning the table the image of the focused detail
does not move with respect to the center of the
crosshair. If not, repeat the operations described
by 2. to 6. until the perfect coincidence of the
center of rotation of the stage with the center of
the crosshair for which the specimen remains in
the center of the crosshair turning the table.
8.
Once centered with the 10x, rotate the nosepiece
to insert in the optical path all the other objectives
and verify the correct centering of the objectives,
acting on the screws of centering the nosepiece
⑥
, to make sure that all the objectives are
perfectly centered with respect to the optical axis.
Fig. 14
①
②
③
Fig. 15
④
⑤
Fig. 16
⑥
⑥
Содержание B-380 Series
Страница 1: ...Model B 383POL B 380 Series INSTRUCTION MANUAL Ver 4 0 2019 ...
Страница 5: ...Pagina 5 Opposite side EYEPIECES DIOPTER ADJUSTMENT RING ANALYZER POLARIZER TENSION ADJUSTMENT RING ...
Страница 22: ...Modello B 383POL Serie B 380 MANUALE DI ISTRUZIONI Ver 4 0 2019 ...
Страница 29: ...Pagina 29 6 Inserirelospinottodell alimentatorenelconnettore posto sul retro del microscopio Fig 5 Fig 5 ...