![Optics11 DELTASENS Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/optics11/deltasens/deltasens_user-manual_1642270023.webp)
23
User Manual
1.5.
Additional pop-up windows
1.5.1.
Sensor select
Figure 14: Sensors select window
Figure 14 shows the Sensors select window. As described in the previous section, it allows to define
new sensors, set their parameters and choose which of the defined sensors the unit should
demodulate
. By double clicking one of the listed sensors the “Sensor select” window will expand
showing the cavity size spectrum as shown in Figure 15.
The extended window allows to set the proper cavity size of each sensor to demodulate its signal.
The nominal cavity size of each sensor is provided on its accompanying datasheet. However,
depending on environmental conditions, the real cavity size might shift from the nominal value and
therefore it should be set by the user after installing the sensor. The nominal value provided in the
Содержание DELTASENS
Страница 1: ...DELTASENS USER MANUAL ...
Страница 14: ...14 optics11 com Figure 6 AFE CTRL submenu Figure 7 Raw time domain signal ...