Step 6 Export Setup File
Below is an image of the Step 6 tab of the spreadsheet including instructions.
INSTRUCTIONS FOR THIS PAGE
Break Point
SMOOTH_CAL1
y =
-0.50
x^2 +
5.90
x +
0.00
1.10
SMOOTH_CAL2
y =
0.00
x^2 +
4.80
x +
0.60
5.10
SMOOTH_CAL3
y =
1.20
x^2 +
-7.44
x +
31.81
MAX
Lasercheck Calibration Curve
0.0
20.0
40.0
60.0
80.0
100.0
120.0
140.0
160.0
0.0
2.0
4.0
6.0
8.0
10.0
12.0
14.0
16.0
Uncalibrated Lasercheck Measurements
A
ct
ua
l R
ou
gh
ne
ss
V
al
ue
s
y = -0.50 x^2 + 5.90 x + 0.00
y = 0.00 x^2 + 4.80 x + 0.60
y = 1.20 x^2 + -7.44 x + 31.81
1. Review the overall calibration curve
fits for all roughness regions. If any part
of the calibration does not visually
appear correct, go back to the tab
corresponding to that region and make
adjustments. You may have to edit other
tabs after that to correct the overall
calibration.
File Transfer Setup File
Manual Input Setup File
RECEIVE_NAME=
ProcessA.set
FILENAME=
ProcessA
FILENAME=
ProcessA
Cal_A1:
-0.50
START_INPUT=
MEAS
Cal_B1:
5.90
RA_UNITS=
MICROINCHES
Cal_C1:
0.00
RA_LIMITS=
15.00
20.00
Cal_A2:
0.00
LOW_ANALOG=
15.00
Cal_B2:
4.80
HIGH_ANALOG=
20.00
Cal_C2:
0.60
RA_THRESHOLD=
12.00
Cal_A3:
1.20
CAL_TYPE=
EXPONENTIAL
Cal_B3:
-7.44
CAL1=
-0.50
5.90
0.00
1.10
Cal_C3:
31.81
CAL2=
0.00
4.80
0.60
5.10
Break_1:
1.10
CAL3=
1.20
-7.44
31.81
MAX
Break_2:
5.10
CAL4=
Threshold:
12.00
CAL5=
RA_UNITS=
MICROINCHES
EOF
INSTRUCTIONS FOR THIS PAGE
1. Click on Push-Buttons Below Labeled "Export Standard Setup File", or "Export
Manual Input Setup File" depending on whether new setup file created will be input
into the Lasercheck electronics by computer file transfer or by manually inputting
values. You may rename any of these files, but they
MUST
end in ".set". Close the
new setup file that is created in Excel. When the dialog box appears asking you
about saving the changes to the file, select "No".
Export "File Transfer"
Setup File
Export "Manual
Input"Setup File