34
Long-term storage
If the sensor is not used over a prolonged period, perform the following
operations.
1. Detach the sensor cable from the transmitter.
2. Drain off the water from the sensor and clean it.
3. Store the unit away from direct sunlight.
As the packing will deteriorate over time, we recommend replacing the
maintenance parts every 3 years. For details, contact the dealer.
Overhaul
14
Specifications
Name
Turbidity Checker
Model
TC-3000-D
TC-3000SP-D (for seawater)
Measurement
range
0 - 3000 FTU
Display
resolution
0.0 - 99.9 FTU: 0.1 FTU
100 - 499 FTU: 1 FTU
500 - 3000 FTU: 5 FTU
Power supply
voltage
24 VDC±10%
Current
consumption
Standard: 15 mA max.
Cleaning: 130 mA max.
Communications
(function)
MODBUS (RTU/ASCII)
Cleaning
system
Both sided swing wiper cleaning system
Calibration
The sensor is designed to be able to measure stably for a prolonged period
of time. In order to maintain the reliability of measurement, however,
calibration should be carried out at least once a year.