Calibration
19
Perform calibration before using the sensor or after cleaning.
Distilled or deionized water is required.
To calibrate SC-U1, press
.
Immerse the sensor in distilled water or ion-exchange water and adjust the
measurement value to the reference value (zero).
If bubbles are present, perform “Manual cleaning” P.30 to remove
bubbles.
The sensor selection display is displayed.
Press
.
4
1
Clean the sensor's main unit and detection window.
Caution
Pressing
automatically retains the measurement value.
3
Allow the sensor to acclimate to the water temperature for
about 5 minutes, and check that there are no bubbles in the
detection window.
Caution
If neither distilled water nor deionized water is available, tap
water without rust/turbidity can be used instead.
2
Immerse the sensor in distilled
water or ion-exchange water.
8
Calibration
Zero calibration