background image

10

11

Operation

5

2

Calibration

Each setting

 Air bubbles on the Detector Windows can be removed by wiping.

Pip

1

Be sure to carry out calibration according to the following procedure 
before using the SC-T4.

4

When pressing the CAL button, "CAL" is 
displayed. "CAL" blinks during calibration.

3

After accustoming the Detector to water temperature for 5 
minutes or so, make sure that air bubbles are not attached 
on the Detector Windows and press       for 2 seconds or 
more.

5

When calibration is completed, the operation panel blips twice and 
the Display turns into normal measurements indication. 

2

Immerse the Detector in distilled water or ion-exchange water. 

1

Clean the Detector and the Detector Windows.

Every time the MODE button          is pressed, the Display is changed into 
each setting. To change the setting value, select the item and then press 
       for 2 seconds or more.

Factory setting and setting range are as shown in the Table below.

Alarm Level

Alarm Timer

Signal Output Response Time

Factory setting

OFF

OFF

OFF

Setting range

 Span-Adjusted value

 120 minutes

 120 seconds

Every time the MODE button       is pressed, the display is changed 
as follows:

 Signal Output High Limit (20mA)

Signal Output Low Limit (4mA)

Signal Output Response Time

Alarm Timer

Alarm Level

Span-Adjustment

0 (ZERO)-Adjustment

Measurements Indication

Ex: The measurement value : 1000 mg/l

Ex: Signal Output Response Time : 10 seconds

Ex: Alarm Timer : 60 minutes

Ex: Alarm Level : 125 mg/l

Содержание SS Checker SC-T4

Страница 1: ...is instruction manual should be kept for future reference such as maintenance Transmitter Instruction Manual 2007 7 59 1362 1 For more information on OPTEX products contact your dealer or visit our we...

Страница 2: ...ture reference such as maintenance Transmitter Instruction Manual 0 1 Table of Contents 1 For safe use 2 External Features 3 Installation Installation Wiring Extension of Detecter Cable 4 Operation Pa...

Страница 3: ...ve gas is in danger of intruding through the piping etc into the Transmitter Do not bind the signal output cable with the power cable or do not put them in the one Cable Clamp Install the transmitter...

Страница 4: ...t On Transmitter Operation panel Display 2 External Features Cover Cable Clamps Key Hook Lock Lever Power Switch Circuit Breakers Terminal Block Operation Buttons Display Terminal Cover Alarm Level In...

Страница 5: ...luding wiring resistance For connection to the Self Checking Relay Output terminal and Alarm Relay Output terminal load resistance should be 240 VAC and 1A Max Although a protective circuit is built i...

Страница 6: ...o carry out Fine Adjustment of Signal Output Low Limit 4mA Press the upper and lower arrow buttons simultaneously to carry out Fine Adjustment of Signal Output High Limit 20mA Move the digit to be ent...

Страница 7: ...d water or ion exchange water 1 Clean the Detector and the Detector Windows Every time the MODE button is pressed the Display is changed into each setting To change the setting value select the item a...

Страница 8: ...s not 1000 use to change the Display into 1000 indication When the display shows 1000 press to terminate the 0 Span Adjustments 8 Press for 2 seconds or more SPAN blinks on the Display Pip Pip Pip 4 A...

Страница 9: ...ayed the value is out of the setting range Alarm Timer can be set for 1 to 120 minutes or OFF Setting unit is 1 minute The Alarm Timer is set for OFF as a factory setting Alarm Timer Setting Ex To set...

Страница 10: ...d lower arrow buttons button increases a numeric value and button decreases the value While pressing buttons the values displayed are continuously changed To cancel the Signal Output Low Limit Setting...

Страница 11: ...Pip Pip Fine Adjustment of Signal Output 6 Fine Adjustment of Lower Limit 4mA and Upper Limit 20mA of Signal Output 4 20mA can be made When both right and left arrow buttons are pressed simultaneously...

Страница 12: ...es measured by the official method and the values indicated by SC T4 at a place where the instrument is installed Indicated values Indicated values 1 2 The following calculating formula is used Span s...

Страница 13: ...indicated on the Display Err 5 is indicated on the Display The Detector is out of order A Detector Cable is not wired properly The Detector is out of order A Detector Cable is not wired properly The D...

Страница 14: ...Power Supply for Sensor 12 VDC Signal Output analog 4 20mA resistance load of 300 or less Self checking Relay Output non voltage C contact capacity 240 VAC 1A resistance load Alarm Relay Output non vo...