Signal Output Response Time Setting

2

Signal Output Response Time can be set for 1 to 120 seconds or OFF.

Setting unit is 1 second

The Signal Output Response Time is set for OFF as a factory setting.

Ex: To set the Signal Output Response Time for 10 seconds ;

3

Enter "10" on the Display.

2

1

Press         for 2 seconds or more, the 
Signal Output Response Time Indicator 
blinks. 

Press         and select the Signal Output 
Response Time. 

Pip

Pip Pip

Signal Output Range Setting

3

Signal Output Range can be set as follows.

"High Limit"

"Low Limit"

 Span-Adjustment value 

 5

    Ex) when the setting value of Span-Adjustment is changed to
          2000mg/l, the Signal Output Setting Range is as follows:
           "High Limit"

"Low Limit"  

 400 ( 

= 

2000 

 5)

The Signal Output Range is default at 4mA for the 0-Adjustment 

value and 20mA for the Span-Adjustment value.

Ex: To set the The Signal Output Range for 100 

 300 mg/l ;

3

Enter "100" on the Display.

2

1

Press         for 2 seconds or more, the 
The Signal Output Low Limit value is 
displayed. 

Press         and select the The Signal 
Output Low Limit. 

Move the digit to be entered using the right and left arrow buttons             
(1 digit       10 digit       100 digit       100 digit). 
At this time, the selected digit blinks.

 

 

Reference 

Enter a numerical value (0, 1, 2, ..., 9) into the selected digit using the 
upper and lower arrow buttons            .         button increases a numeric 
value and          button decreases the value.
While pressing           buttons, the values displayed are continuously 
changed.
To cancel the Signal Output Low Limit Setting, press         .

(1) Signal Output Low Limit Setting

Pip

 "-Lo-" and "the setting value" comes on 

alternatively.

Pip Pip

CAUTION:

 The resolution of the Analog Signal Output change depending on the setting value of 

the Span-Adjustment. 

 The resolution of Analog Signal Output is 0.5 mg/l as a factory setting. When it is 

necessary to get 0.1mg/L as the resolution of Analog Signal Output, set the Span-
Adjustment to "1000mg/l" and the Signal Output Range to "200mg/l". 

4

Press         to terminate the Signal Output Response 
Time Setting.

Move the digit to be entered using the right and left arrow buttons             
(1 digit       10 digit       100 digit). 
At this time, the selected digit blinks.

 

 

 

Reference 

Enter a numerical value (0, 1, 2, ..., 9) into the selected digit using the 
upper and lower arrow buttons            .        button increases a numeric 
value and          button decreases the value.
While pressing           buttons, the values displayed are continuously 
changed.

Reference:

To cut off the Signal Output Response Time, enter "0". Then "OFF"appears on the 

Display. Next, press         .

To cancel the Signal Output Response Time  Setting, press         .

CAUTION:

 When Err6 is displayed, the value is out of the setting range.

 

To change the Display for the Measurements Indication mode, press        
again.

 

 

To change the Display for the Measurements Indication mode, press        
again.

Содержание SS Checker SC-T4

Страница 1: ...is instruction manual should be kept for future reference such as maintenance Transmitter Instruction Manual 2007 7 59 1362 1 For more information on OPTEX products contact your dealer or visit our we...

Страница 2: ...ture reference such as maintenance Transmitter Instruction Manual 0 1 Table of Contents 1 For safe use 2 External Features 3 Installation Installation Wiring Extension of Detecter Cable 4 Operation Pa...

Страница 3: ...ve gas is in danger of intruding through the piping etc into the Transmitter Do not bind the signal output cable with the power cable or do not put them in the one Cable Clamp Install the transmitter...

Страница 4: ...t On Transmitter Operation panel Display 2 External Features Cover Cable Clamps Key Hook Lock Lever Power Switch Circuit Breakers Terminal Block Operation Buttons Display Terminal Cover Alarm Level In...

Страница 5: ...luding wiring resistance For connection to the Self Checking Relay Output terminal and Alarm Relay Output terminal load resistance should be 240 VAC and 1A Max Although a protective circuit is built i...

Страница 6: ...o carry out Fine Adjustment of Signal Output Low Limit 4mA Press the upper and lower arrow buttons simultaneously to carry out Fine Adjustment of Signal Output High Limit 20mA Move the digit to be ent...

Страница 7: ...d water or ion exchange water 1 Clean the Detector and the Detector Windows Every time the MODE button is pressed the Display is changed into each setting To change the setting value select the item a...

Страница 8: ...s not 1000 use to change the Display into 1000 indication When the display shows 1000 press to terminate the 0 Span Adjustments 8 Press for 2 seconds or more SPAN blinks on the Display Pip Pip Pip 4 A...

Страница 9: ...ayed the value is out of the setting range Alarm Timer can be set for 1 to 120 minutes or OFF Setting unit is 1 minute The Alarm Timer is set for OFF as a factory setting Alarm Timer Setting Ex To set...

Страница 10: ...d lower arrow buttons button increases a numeric value and button decreases the value While pressing buttons the values displayed are continuously changed To cancel the Signal Output Low Limit Setting...

Страница 11: ...Pip Pip Fine Adjustment of Signal Output 6 Fine Adjustment of Lower Limit 4mA and Upper Limit 20mA of Signal Output 4 20mA can be made When both right and left arrow buttons are pressed simultaneously...

Страница 12: ...es measured by the official method and the values indicated by SC T4 at a place where the instrument is installed Indicated values Indicated values 1 2 The following calculating formula is used Span s...

Страница 13: ...indicated on the Display Err 5 is indicated on the Display The Detector is out of order A Detector Cable is not wired properly The Detector is out of order A Detector Cable is not wired properly The D...

Страница 14: ...Power Supply for Sensor 12 VDC Signal Output analog 4 20mA resistance load of 300 or less Self checking Relay Output non voltage C contact capacity 240 VAC 1A resistance load Alarm Relay Output non vo...

Отзывы: