OPTEX FA CS-30TAC Скачать руководство пользователя страница 13

13

 TECH

 E1.00

   150

   150

   150

   160

   160

 E0.90

 E

 E1.00

 E

 E0.95

Emissivity rate (ε)

The emissivity rate is the rate of energy emitted from the surface of an object. Every object has 

a unique emissivity rate which is variable according to the surface condition and temperature of 

the object. This product allows for setting a desired emissivity rate, which can enable even more 

precise measurement by adjusting the emissivity rate according to that of the target.

An object with low emissivity rate (e.g. a shiny metallic object) reflects the surrounding 

temperature since it is highly reflective. If the surrounding objects have greatly different 

temperature from that of the main unit, their temperatures are reflected resulting in incorrect 

measurement. Therefore it is necessary to block out such effect.

The maximum emissivity rate is normally 1.00, but this unit is designed to accept up to 1.20 for 

practical convenience.

(1) Aim the sensor head to the target.

 

Confirm the target is sufficiently larger than the field of view.

(2) In the Setting Operation select the Adjustment Setting, and then press Enter button.

(3) Select the TEACH Adjustment, and then press the ENTER button.

(4) The current measurement value is displayed. 

 

Input the temperature of the target, and then press the ENTER button.

(5) Confirm that the measurement value and emissivity rate have been changed.

Make sure to perform Save Settings.

(1) In the Setting Operation select the Adjustment Setting, 

and then press Enter button.

(2) Select the Emissivity Rate Adjustment, and then press 

the ENTER button.

(3) Input the emissivity rate, and then press the ENTER button.

(4) Confirm that the measurement value and emissivity rate 

have been changed.

Make sure to perform Save Settings.

TEACH Adjustment

Emissivity Rate Adjustment

 Err1

 E

 

Error indication

If the emissivity rate calculated automatically in the TEACH Adjustment is out of the setting 

range(0.1 to 1.2), an error(Err1) is displayed.

In this case, TECH Adjustment is cancelled. Perform Emissivity Rate Adjustment.

Input the emissivity rate directly.

Adjust the measurement value to the temperature value of the target. 

The emissivity ratio is calculated automatically. 

Содержание CS-30TAC

Страница 1: ...S 30TAC CS 40TAC CS 30TAC HT CS 40TAC HT 91 Chudoji Awata cho Shimogyo ku Kyoto 600 8815 JAPAN TEL 81 75 325 1314 FAX 81 75 325 2936 OPTEX FA CO LTD Non Contact Thermometer Printed in JAPAN 0807230 20...

Страница 2: ...to convert the infrared energy emitted from the surface of an object into temperature This thermometer measures the surface temperature of solid and liquid without contacting them The temperature of...

Страница 3: ...off DO NOT TOUCH THE LENS Do not touch the lens with hard or sharp objects Do not insert foreign objects into the light receiving part Otherwise incorrect measurement will occur KEEP THE THERMOMETER A...

Страница 4: ...Emissivity rate adjustment 0 1 to 1 2 Power supply 12 to 24 VDC 10 Consumption 120 mA Max load 80 mA Eco mode Ambient temperature Sensor head 0 to 180 C Amplifier 0 to 65 C Sensor head 0 to 100 C Amp...

Страница 5: ...ens surface 7 M12 P1 0 10 30 14 Sensor head Sensor cable 4 3000 MAX Cable cover Serial No buttons ENT button Output cable 6 2000 MAX Cable No Main display Amplifier Sub display 64 52 35 39 31 mm DIN r...

Страница 6: ...2 Do not wire 15 Orange Black 2 Connect to GND 16 4 Shielded cable Output cable 1 Connect to the 4 20 mA input of an analog device Analog output allowable load 250 and analog output impedance 47 2 Ex...

Страница 7: ...Insert the connector to the cable cover The metal plate of connector should be placed onto the clasp in the cable cover Sensor cable 5 15 5 Cable tie of approx 2 mm width Guide wire 1 2 3 4 mm Connect...

Страница 8: ...al screw is M12 P1 0 Fix securely into the hole of 12 mm or more using the attached hexagon nut The optional mounting fitting can help you adjust the angle easily Mounting Mount the sensor head perpen...

Страница 9: ...w shown above When measuring a high temperature target keep as much distance from it as possible within the range of the field of view Otherwise it may cause incorrect measurement When using the optio...

Страница 10: ...ation Press 3 sec or more Setting Operation Make sure to perform Save Settings If you cancel in Save Settings all settings entered are not saved 8 8 8 8 8 8 8 8 3 To confirm or change the settings pre...

Страница 11: ...d alarm ON OFF can be set The Unit has 3 Alarm Output Modes as below NORMAL Mode Alarm output is kept on as long as the value is above or below the limit value DELAY Mode Alarm output turns on when a...

Страница 12: ...e bank 1 to 4 to change the setting Settings of Response Time Adjustment Analog Output Alarm Output Trigger Input and Display Off can be saved in each bank 1 In the Setting Operation select the Bank S...

Страница 13: ...ently larger than the field of view 2 In the Setting Operation select the Adjustment Setting and then press Enter button 3 Select the TEACH Adjustment and then press the ENTER button 4 The current mea...

Страница 14: ...not change the emissivity rate after setting in 2 Point Adjustment otherwise the adjusted value will be changed 7 Aim the sensor head to the target Confirm the target is sufficiently larger than the f...

Страница 15: ...40 L 0 3 L Lower limit value is displayed Input the temperature L lower limit value and then press Enter button Make sure to perform Save Settings The minimum width of output range is 100 degrees 0 t...

Страница 16: ...on when the value exceeds or falls below the limit value and is kept for a specific time NORMAL Mode Alarm output is kept on as long as the value is above or below the limit value DELAY Mode Alarm ou...

Страница 17: ...value and then press the ENTER button 6 Select NORMAL Mode and then press the ENTER button Make sure to perform Save Settings 6 Select DELAY Mode and then press the ENTER button 7 The DELAY Time is d...

Страница 18: ...og output types as below HOLD MIN MIN HOLD P P P P HOLD SAMP SAMPLE Setup of Functions MIN HOLD The minimum value between the triggers is output HOLD MAX MAX P P HOLD The P P value maximum minus minim...

Страница 19: ...hen press Enter button Make sure to perform Save Settings WAVE Mode HOLD MAX HOLD MIN WIMT 200 WLIM 150 WIMT 200 WLIM 150 Analog Output is controlled by External Trigger Analog Output is controlled by...

Страница 20: ...ing to the analog output of MAX HOLD set by the trigger input alarm is output while the value is outside the range between H and L ECO OFF 1 In the Setting Operation select the Display Off Setting and...

Страница 21: ...gger mode MAX MIN MAX MIN MAX MIN MAX MIN Trigger temperature Display Off Time to turn off s s s s OFF Note of the settings Save or cancel the settings entered in the Setting Operation Switch other ba...

Страница 22: ...s is dirty remove the dust using a blower for cleaning lens For stubborn dirt apply a small amount of ethyl alcohol to a cotton swab or special lens cleaning cloth and gently wipe off the dirt Amplifi...

Отзывы: