F-2200 Series Vortex Flow Meter Installation and Operation Guide • Revised 12/08

21 

REPORTED 

PROBLEM

POSSIBLE 

SOLUTIONS

A non-zero fl ow indi-
cated when no actual 
fl ow is in the pipe.

• Mains interference due to improper Earth ground connection. The protective 

Earth PE terminal should be properly grounded (see page 17).

• Excessive mechanical vibration in the pipe. If so, support the pipeline near the 

meter perpendicular to both the axis of the pipe and the axis of the shedder 
bar.

• This problem may also be solved by reducing the factory set gain. (Contact 

ONICON for assistance.) 

NOTE: By reducing the gain, the minimum measurable fl ow rate will go up 
by the factor which is approximately equal to the square root of the gains 
(odd gains new gain). If the minimum fl ow with reduced gain is above the 
minimum fl ow which is required to be measured, then reducing the gain is 
not the permanent solution. Then the installaion should be corrected and 
also the vibrations should be eliminated.

“CHECK INST.” error 
is displayed when no 
fl ow is in the pipe.

The display should normally indicate 0.0 fl ow rate, LOW FLOW or LOW SIGNAL 
when there is no fl ow in the pipe. The additional CHECK INSTALL error (fl ow 
rate = 0.0 or some steady or fl uctuating value) is an indication of:

• Improper/inadequate earthing
• Excessive pipe vibration

Flow rate inidcated is 
0.0 even with fl ow in 
the pipe.

• The vortex sensor cable is disconnected or is not properly connected.
• Flow sensor is faulty. Contact ONICON for assistance.

The fl ow indicated 
responds to changes 
in the fl ow but the 
indicated value does 
not correspond to the 
actual fl ow rate. Also 
“CHECK INST.” Error 
may appear intermit-
tently.

• The meter is not properly centered on the pipeline. The axis of the meter bore 

should be aligned with that of the pipe.  

• Gaskets at the meter are protruding into the pipe bore. The gaskets must not 

project into the effective cross-section of the pipe.  

• Irregularities on the surface of the pipe bore. The pipe bore should be free from 

irregularities at the welded joints, dirt, deposits and excessive surface rough-
ness.  

• The Vortex signal is distorted due to a bi-phase medium. Bi-phase media are 

not permitted. Use a moisture separator for wet steam applications to remove 
the moisture droplets from the steam. Use suitable fi lters in gas applications to 
remove solid particles from the fl owing gas.  • Incorrect angular position of the 
meter. Refer to Section 3.1.1 for the allowable mounting positions.  

• Insuffi cient upstream/downstream pipe lengths. Check that the upstream/

downstream pipe lengths are of the correct minimum length as given in 
Section 3.1.3.  

• Error in meter factor K-Factor programming. Contact ONICON for assistance.  
• Check that the fl ow direction and the direction arrow on the meter body agree. 

4.5 TROUBLESHOOTING 

HINTS

Содержание Vortex Flow Meter F-2200 Series

Страница 1: ...ies Vortex Flow Meter Installation and Operation Guide For software version CF 4 7 and higher 1500 North Belcher Road Clearwater FL 33765 Tel 727 447 6140 Fax 727 442 5699 www onicon com sales onicon com 12 15 08 ...

Страница 2: ...se of this product ONICON assumes no responsibility for any such harm owing to any defects in the product which were not predictable or for any indirect damages Safety Precautions The following general safety precautions must be observed during all phases of installation operation service and repair of this product Failure to comply with these precautions or with specific WARNINGS given elsewhere ...

Страница 3: ...n Guide Addendum December 12 2008 CAUTION Insulating the entire neck of the flow meter will increase heat transfer to the electronics enclosure and may in some cases cause premature failure of the electronics The drawing below illustrates the proper method for insulating the flow meter F 2000 SERIES STEAM FLOW METER WITH INSULATION DETAIL ...

Страница 4: ...pipe 12 3 1 3 Straight unimpeded inlet and outlet runs 13 3 1 4 Minimizing pipe vibration 14 3 1 5 Locating the meter in a pipeline that runs parallel to a wall 14 3 1 6 State of medium 14 3 1 6 1 Liquid Applications 14 3 1 6 2 Steam or compressed gas applications 14 3 2 Mechanical Installation 15 3 2 1 Orient the display for convenient viewing 15 3 2 1 1 Liquid applications 15 3 2 1 2 Steam or co...

Страница 5: ...F 2200 Series Vortex Flow Meter Installation and Operation Guide Revised 12 08 6 ...

Страница 6: ...ocity of the medium When provided with optional temperature and or pressure compensation the F 2000 series flow meters are capable of calculating the required density corrections necessary to provide direct mass flow measurement of liquids steam or gases 1 3 FEATURES AND SPECIFICATIONS CONSTRUCTION Meter body Casting 316 stainless steel for diameters up to 4 304 stainless steel for 6 and 8 diamete...

Страница 7: ...pad and magnetic pin programming OUTPUT SIGNALS Rate 4 20mA loop powered Supply voltage range 14 36 VDC Minimum load resistance 100 ohms Maximum load resistance is calculated as follows Max load Vs 12V Where Vs supply voltage 0 020A Total Scaled output pulse programmable Optically isolated open collector 0 5 hz maximum Rating 5 36 VDC 100mA maximum SUPPLY VOLTAGE 14 36 VDC PROCESS CONNECTIONS Stan...

Страница 8: ...ed for use in industrial environments that are free of corrosive liquids fumes and excessive vibration Do not expose the flow meter electronics enclosure to direct sunlight Install a sunshade if necessary Do not expose the flow meter to intense vibration If necessary provide additional support to the pipeline at the meter location to minimize vibration The rotating design of the electronics enclos...

Страница 9: ...nd notify the freight carrier and ONICON immediately if any items arrive damaged 2 1 ENSURING THAT YOU HAVE RECEIVED EVERYTHING Standard Documentation Each ONICON flow meter is serialized and supplied with a comprehensive documentation package specifically prepared for this meter It includes the following items Please notify ONICON immediately if any discrepancies are found Flow Meter The F 2000 V...

Страница 10: ...the point in the piping system where the flow meter will be installed will ensure accurate and reliable operation When selecting an installation site consider the criteria contained in Section 1 5 WORKING ENVIRONMENT as well as the following 3 1 1 Flow Direction And Meter Position The flow must always be in the direction of the arrow located on the meter body This will orient the bluff side of the...

Страница 11: ...pe ID Difference in 1 1 05 0 016 1 1 2 1 61 0 016 2 2 07 0 024 3 3 07 0 024 4 4 03 0 024 6 6 07 0 031 8 7 98 0 039 Ensure that the bore of the locating pipes are smooth and without deposits scaling or welding beads 3 1 2 Maximum Allowable Difference Between Inside Diameter Of Meter Body And The Connecting Pipe The table below provides the maximum allowable difference between the diameter of the fl...

Страница 12: ...lve upstream the inlet length is 25D instead of 50D The minimum inlet pipe length including the flow straightener must always be at least 12D 3 1 4 Minimizing Pipe Vibration Pipe vibration caused for example by the action of pumps valves etc will distort flow measurements particularly at low flow velocities To minimize the effects of vibration support the pipeline on both sides of the meter in a d...

Страница 13: ...e liquid are not permitted 3 1 6 1 Liquid Applications When operating with liquids the meter requires a minimum downstream pressure to prevent cavitation The formula used to determine the minimum downstream pressure is as follows Pds bar_g 2 9 DP 1 3 Ps 1 013 Where DP pressure drop of the meter in Bar as determined by the ONICON sizing program and Ps saturation pressure in Bar at the operating tem...

Страница 14: ...er is installed 3 2 1 1 Rotating the electronics enclosure The electronics enclosure is attached to the meter body with 4 5mm Allen head screws To rotate the enclosure you must first remove the screws and then carefully rotate the enclosure as needed The interconnecting stem assembly is a conduit for wires that connect the sensors to the circuitry contained in the enclosure Care must be taken to p...

Страница 15: ... certain that the interconnecting cable is not damaged in the process 3 2 2 Flanged Type Connection To Ansi B16 5 Schedule 40 Meter sizes 3 8 1 2 3 4 1 1 1 2 2 3 4 6 8 Pipe flanges To ANSI 300 standard 600 optional Gaskets are supplied with flanged units Center the flow meter by sight Check the flange connections for leak tightness after the flow meter installation CAUTION Do not remove enclosure ...

Страница 16: ...ured Flow meters supplied with this option will display the medium temperature and deliver an output signal that is temperature compensated for mass and or normalized flow 3 3 WIRING TERMINATIONS The standard F 2000 series vortex meter is a 2 wire loop powered device Wiring connections are shown below CAUTION Do not remove enclosure covers in any wet environment Keep dirt and debris out of electro...

Страница 17: ...he display indicates the current measured value The second line of the display indicates the programmed units of measurement There are two options for displaying data in the measurement mode The display may be set to automatically scroll through each menu page or it may be set to manually step through the menu pages using the up arrow key to advance each page Displays operating in the cyclic mode ...

Страница 18: ... 3 corresponding Hall Effect switches that perform the same functions The Hall Effect switches may be activated without removing the display cover using the magnet provided with the meter Place the magnet against the cover in proximity to the Hall Effect switch to activate IMPORTANT NOTE Do not attempt to enter the program mode without first contacting ONICON service CAUTION Do not remove enclosur...

Страница 19: ...ow rate However accuracy of measurement may suffer HIGH FLOW Actual flow rate higher than specified range Corrective action depends on application pro cess If the flow rate exceeds the maximum value it may damage the sensor physically INV CONFIG Configuration data in non volatile memory is not valid Contact ONICON ISO FAIL Checked during power on only Sensor isolation has failed Contact ONICON AMP...

Страница 20: ...e pipe vibration Flow rate inidcated is 0 0 even with flow in the pipe The vortex sensor cable is disconnected or is not properly connected Flow sensor is faulty Contact ONICON for assistance The flow indicated responds to changes in the flow but the indicated value does not correspond to the actual flow rate Also CHECK INST Error may appear intermit tently The meter is not properly centered on th...

Страница 21: ...ener significantly reduces the upstream straight pipe length requirement for ONICON Vortex Flow Meters The size of the straightener should always match the meter size as opposed to the original pipe size The flow straightener is made of 304 A 351 CF8 stainless steel SIZE DIM A DIM B DIM C DIM D NUMBER OF HOLES 2 3 93 1 0625 28 2 14 35 3 5 31 1 0625 43 3 24 35 4 6 26 1 0625 55 4 22 35 6 8 50 1 0625...

Страница 22: ...BETWEEN FLOW METER AND STRAIGHTENER SINGLE 90 12 DIA 10 DIA TEE 12 DIA 10 DIA RDCR EXPNDR 12 DIA 10 DIA TWO 90 S SAME PLANE 17 DIA 15 DIA BALL GATE VALVE FULLY OPEN 17 DIA 15 DIA TWO 90 S OUT OF PLANE 22 DIA 20 DIA CONTROL VALVE 27 DIA 25 DIA PRV 27 DIA 25 DIA NOTES 1 Consult ONICON for meter size and applicable meter pipe run for each application Install according to manufacturer s recommendation...

Отзывы: