F-5000 Software Manual 10/16 - 2032 / 107050
Page 25
Zero CAL-CHECK™ Performed on:
September 23 2016
10:30:29
Firmware version:
F-5500 V3.6
Serial Number:
N00003
Zero CAL-CHECK™ Results:
ZERO CAL-CHECK PASS
Zero CAL-CHECK™:
0.32
Test Temperature
73.1 F
Tag #/Meter Location:
Test performed by:
Additional Comments:
* Sensor free of residue or film
* Zero Stability of the meter
* Thermal Conductivity (heat transfer) repeatability of the sensor
Zero CAL-CHECK™ is a calibration routine that validates the flow meter's calibration accuracy by testing the following:
At the conclusion of the test, the meter will display a pass/fail message and the Zero CAL-CHECK™ data.
A "pass" result confirms the meter is measuring accurately.
CALIBRATION VALIDATION
Zero CAL-CHECK™ CERTIFICATE
Zero CAL-CHECK
®
Certificate
The Zero CAL-CHECK
®
Certificate function displays the latest certification. When
performing a Zero CAL-CHECK
®
test, all the data is logged into a log file with all
pertinent data, including the serial number. A laptop or PC can be used to perform the
Zero CAL-CHECK
®
test on the F-5000 meter. When a Zero CAL-CHECK
®
certificate
is requested, the program will search the log file for the specific serial number and will
display only the last check performed.
Fig. 3.19: Zero CAL-CHECK
®
Certificate
Zero CAL-CHECK
®
Log
The "View Log" button allows the operator to view a log of previous Zero CAL-CHECK
®
tests that have been executed on the meter.
Fig. 3.20: Zero CAL-CHECK
®
Log
F-5000 View
OPERA
TION
Operation: Zero CAL-CHECK
®
Certificate
Zero CAL-
CHECK
®
Certificate
Zero CAL-
CHECK
®
Log