1500 North Belcher Road, Clearwater, FL 33765 • Tel (727) 447-6140 • Fax (727) 442-5699 • [email protected]
F-3200 Flow Meter Manual 04/13 - 0671-12 / 18325
Page 11
SECTION 2.0: UNPACKING
The F-3200 is generally shipped in one package unless optional hardware or equipment is ordered.
Notify the freight carrier (all products are shipped insured) and ONICON if any items are damaged in
transit.
2.1 CHECKING THAT YOU HAVE RECEIVED EVERYTHING
• Standard Documentation
Enclosed with each F-3200 is a comprehensive documentation package that includes
the following items:
The F-3200 Installation and Operation Guide
Certifiat of Calibration
Please notify ONICON if any of these items are missing.
METER DATA
CALIBRATION of PRIMARY FLOW ELEMENT
Meter Tag:
Model:
Serial Number:
220177
Component S/N's:
04N1420 38M6901
Meter Size:
3"
Max. Operating Pressure:
225 psi
Max. Operating Temperature:
212 °F
Connections:
ANSI 150# Class Flanges
Peripheral device serial number
220177
Analog Flow Range: 4-20 mA = 0 to
230 Gal/Min
Full Scale Frequency Output:
200.00 Hz
Scaled Pulse Output:
1 pulse =
100 Gallons
GL
Date:
4/12/2011
5/25/2011
This meter was manufactured for ONICON Incorporated.
The original manufacturer certifies that this flow meter was
calibrated against standards that are traceable to
SIT, Italy.
1500 N Belcher Road, Clearwater, Florida 33765 Tel (727)447-6140 Fax (727)442-5699
Programmed By:
FACTORY PROGRAMMED OUTPUT SIGNALS
(Performed at ONICON Factory; can be reprogrammed in the field)
Primary Calibration Date:
F-3103-111
FLOW METER
CERTIFICATE OF CALIBRATION
CALIBRATION & CONFIGURATION DATA for F-3000 SERIES MAGNETIC FLOW METERS
CT-WH
-1.0951
Water
Medium:
Ka factor:
F-3200 Series
In-line Electromagnetic Flow Meter