Troubleshooting
9
90
What it means
The sample was applied before the meter was ready.
What to do
Repeat the test with a new test strip. Apply a whole blood sample, control, or
linearity solution only after
Apply Blood
or
Apply Solution
appears on the display.
If this message continues to appear, contact Customer Service and/or LifeScan®
sales representative.
ERROR 3
Meter was not
ready. Retest
with a new strip.