Detailed information
about your system
9
136
User Accuracy
A study evaluating glucose values from fingertip capillary
blood samples obtained by 165 lay persons showed the
following results:
•
100% within ±0.83 mmol/L of the medical laboratory
values at glucose concentrations below 5.55 mmol/L,
and 97.7% within ±15% of the medical, laboratory
values at glucose concentrations at or above
5.55 mmol/L
98.2% of the total number of samples were within
±0.83 mmol/L or ±15% of the medical laboratory values
Guarantee
LifeScan guarantees that the OneTouch Select® Plus
Meter will be free of defects in material and workmanship
for three years, valid from the date of purchase. The
guarantee extends only to the original purchaser and is
not transferable.