Detailed information
about your system
9
131
System Accuracy
Diabetes experts have suggested that glucose meters
should agree within 0.83 mmol/L of a laboratory
method when the glucose concentration is lower than
5.55 mmol/L, and within 15% of a laboratory method
when the glucose concentration is 5.55 mmol/L or higher.
Samples from 100 patients were tested using both the
OneTouch Select® Plus System and the YSI 2300 Glucose
Analyser laboratory instrument.
System Accuracy Results for Glucose
Concentrations <5.55 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within
±0.28 mmol/L
Within
±0.56 mmol/L
Within
±0.83 mmol/L
72.0%
(121/168)
94.6%
(159/168)
98.2%
(165/168)