Timing Charts
121
ZX2 User’s Manual
INTRODUCTION
BASIC
SETUP
PREPARATION
FOR
MEASUREMENT
TROUBLE-
SHOOTING
INDEX
CONTENTS
SETTING
TRANSITION
CHARTS
FLOW OF
OPERATION
MAIN
APPLICATIONS
& SETTING
METHODS
DETAILED
SETTINGS
SPECIFI-
CATIONS
Height
Double
Sheet
Detection
Positioning
Eccentricity
and Surface
Deflection
Steps
and
Warpage
Thickness
• When the zero reset memory setting is ON
1
Zero reset
execution
Turn OFF after 4 ms to 1 s zero reset input turns ON.
The zero reset is executed, and measurement is resumed within
400 ms.
Acquire the measurement results after the preset response time has
elapsed.
2
Zero reset
cancellation
Turn OFF after zero reset input turns ON for 3 s or more.
The zero reset is canceled, and measurement is resumed within
400 ms.
Acquire the measurement results after the preset response time has
elapsed.
OFF
ON
3 s
Within
400 ms
Within
400 ms
Zero reset
execution
Zero reset
cancellation
Zero reset input
Measurement
Zero reset
Output
or more
Result after
zero reset
cancellation
Result after
zero reset
execution
4 ms
to 1 s
Содержание ZX2 Series