
71
ZJ-SD
User’s Manual
S
e
ct
ion 5
DETAILE
D
S
E
TTINGS
Section 5
Linear Output
This section describes how to set the output range, using an example of current output
with a range with the following conversions: 5 kV to 4 mA and 15 kV to 20 mA. Change
the values in this example for voltage output as necessary.
1.
Turn OFF the Amplifier Unit.
2.
Move the current/voltage switch to current
output. The switch is located on the bottom of
the Amplifier Unit.
The default setting is voltage output.
Moving to the FUN mode and SPCL
3.
Turn ON the Amplifier Unit and move the
mode switch to FUN.
4.
Use the LEFT and RIGHT keys to display
[SPCL] on the main display.
Moving to FOCUS
5.
Press the UP or DOWN key.
The sub-display flashes.
6.
Use the UP and DOWN keys to display [SET]
or [ALL].
7.
Press the ENT key.
20
4
0
20
5
15
4
(A)
(B)
0
20
Output current
(mA)
Measured value
(kV)
Output current
(mA)
Measured value
(kV)
Current output
Voltage output
operation2.book Page 71 Wednesday, November 29, 2006 2:56 PM