1. Open flames or smoking during oxygen therapy are dangerous and may result in fire, facial burns, or 
death. Do not allow smoking or open flames within 2m of the device or any oxygen carrying accesso-
ries, such as an attached cannula.
2. Do not use oil, grease or petroleum-based products on or near the device as the use of such products 
may damage the electronic components of the device.
3. For proper operation, the device requires unobstructed ventilation.The device should be located as 
to avoid pollutants or flames.
4. Do not plug or unplug the power cord when there is water on your hand. Otherwise it may cause 
electric shock.
5. Do not use alcohol, solvents, polishes or any oily substances on the device. Otherwise, the output 
oxygen may be contaminated, causing injury or physical discomfort.
6. When placing the oxygen concentrator, please keep a distance of 16cm from walls, curtains or other 
objects, otherwise it may affect the air exchange of the air exhaustion of oxygen concentrator.
7. The device should not be used adjacent to or stacked with other equipment because it could result in 

improper operation .
8. Do not use the oxygen concentrator if either the plug or power cord is damaged . Do not use 
extension cords or electrical adapters.
9. Do not attempt to clean the concentrator while it is plugged into an electrical outlet.
10. The oxygen concentrator regularly emits a "ta..pu" exhaust sound when the oxygen concentrator 
is operating (about 7 seconds).
11. The manufacturer recommends that the oxygen concentrator runs for no less than 30 minutes 
each time when turned on. Do not turn on and off the oxygen concentrator frequently. It is suggested 
to turn on/ turn off after 3 to 5 minutes each time, to avoid affecting the life of the compressor.
12. The use of accessories or service equipment not specified for use with the device may impair its 
performance.
13. Please do not adjust the flow rate above 5L/min, otherwise the oxygen purity may not reach more 
than 90%.
14. When the oxygen purity is abnormal, the user should contact the customer support for consulta-
tion (see the back cover).
15. If you begin to feel ill or are experiencing discomfort while using the oxygen concentrator,  please  
go to the doctor immediately.
16. When the user is a child or a person with limited mobility, the device needs to be used under the 
supervision of the custodian.
17. If you have plan not to use the oxygen concentrator for several days, please clean and dry the 
humidification bottle for storage.
18. Molecular sieve is a consumable, it will absorb moisture from the air during use and storage, and 
its performance will decay, resulting in a decrease in oxygen purity. Before the first use, if the oxygen 
purity is lower than 90%, please contact the customer support for consultation (see the back cover).
19. Do not leave the nasal oxygen cannula on bed covering or chair cushion, if the oxygen concentra-
tor is turned on but not in use, the oxygen will make the materials flammable. Turn the device off 
when it is not in use.
20. The meaning of symbols.

Stacking layer limit

Manufacturer

This way up

Alternating current 

No smoking  

Keep dry

Class Ⅱ equipment

Consult instructions 

for use

Symbol               Meaning

Symbol               Meaning

Symbol               Meaning

Indicates a warning 

or note

Temperature limit

Humidity limitation

Atmospheric

pressure limitation

Содержание Y-506W

Страница 1: ...manual Thank you for purchasing the OMRON oxygen concentrator In order to use the device safely please read the complete manual carefully before using it for the rst time Keep this instruction manual...

Страница 2: ......

Страница 3: ...scope of application of the product Product features Know your device Operation steps Maintenance instructions XIII Schematic diagram of gas circuit and electric circuit XIV Troubleshooting XV Compone...

Страница 4: ...customer shall not disassemble and repair it privately If abnormal phenomena such as the alarm is found please contact the customer support for consultation see the back cover SAFETY INSTRUCTIONS 1 Op...

Страница 5: ...ou begin to feel ill or are experiencing discomfort while using the oxygen concentrator please go to the doctor immediately 16 When the user is a child or a person with limited mobility the device nee...

Страница 6: ...e stored in an environment like this No strong sunlight no corrosive odor well ventilated and clean Avoid violent vibration inverted or horizontal placement during transportation THE SCOPE OF APPLICAT...

Страница 7: ...tor oxygen purity better and understand the use condition of patients oxygen purity 82 only the green light of the operation indicator is on 50 oxygen purity 82 yellow light is on oxygen purity 50 the...

Страница 8: ...ce should not be used with ammable anesthetic gas mixed with air or ammable anesthetic gas mixed with oxygen or nitrous oxide 5 years except fragile parts and consumables Maximum outlet pressure Compr...

Страница 9: ...omizer Primary ltration 06 Rear view of the whole device Side view of the whole device Power switch Including fuse Nebulization outlet Secondary lter Humidi cation bottle Device power jack Primary lte...

Страница 10: ...he device will stop oxygen producing and enter the standby state after the timing is completed Set the timing for 8 hours and press the Timing button again to enter the continu ous oxygen therapy mode...

Страница 11: ...oxygen outlet of the humidi cation bottle was blocked the pressure relief valve of the humidi cation bottle will operate by itself At this time oxygen will be ejected from the pressure relief valve bu...

Страница 12: ...the network power socket and the power cord connector is inserted into the device power jack Then the device can get power supply 9 Primary lter Contains lter sponge for air ltration 10 Secondary lte...

Страница 13: ...cation bottle and then put the nasal oxygen cannula on the user s ears inserting the nasal congestion in the user s nose as shown in the gures below Please follow the doctor s advice for the time and...

Страница 14: ...zation kit 2 When the nebulization function is used the oxygen purity will be reduced and the device may alarm and the alarm will be automatically released when the oxygen ow is reduced It is not a ma...

Страница 15: ...the device and the display screen Most cleaners must be diluted before use 2 Filters The cleaning and the replacement of the lters protect the compressor and molecular sieve and Symbol State Status In...

Страница 16: ...sulting about lter please contact customer support see the back cover 3 Humidi cation bottle The oxygen concentrator is equipped with a humidi cation bottle and it is recommended to clean it each time...

Страница 17: ...new fuse Figure 3 reinstall the fuse holder Figure 4 Figure 4 Figure 3 Figure 2 Figure 1 Air Filter Compressor Oxygen and nitrogen separation Adsorption tower Humidi cation bottle Oxygen outlet Oxyge...

Страница 18: ...ss the ON OFF button of the operation panel as required 3 Turn on the power switch and restart the device 4 Replace the fuse Firstly cut o the power supply and carry out by professional maintenance pe...

Страница 19: ...3 If the phenomenon continues please contact the customer support for consultation see the back cover 7 The inside of the nasal oxygen cannula has a lot of fog or water droplets the method of removing...

Страница 20: ...it has been deformed during cleaning and disinfection 2 Replace the disposable kit 3 The cup mouth of the nebulization cup is blocked 3 Clean or replace the nebulization cup 4 Excessive addition of li...

Страница 21: ...with the corresponding national laws and regulations It is recommended to consult the environmental protection department or the Omron customer support when dealing with the corresponding waste and r...

Страница 22: ...n the manual 2 Portable and mobile radio frequency communication equipment may a ect the normal operation of this product 3 The maximum length of the power cord of this product is 2m 4 In addition to...

Страница 23: ...le for use in all establishments including domestic establishments and those directly connected to the public low voltage network that supplies buildings used for domestic purposes Harmonic Emissions...

Страница 24: ...the test voltage is applied Floors should be wood concrete or ceramic tile If oors are covered with synthetic material the relative humidity should be at least 30 Mains power quality should be that o...

Страница 25: ...he eld strengths should be less than 3V m Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distanc...

Страница 26: ...e by maintaining a minimum distance between portable and mobile RF communications equipment trans mitters and this concentrator as recommended below according to the maximum output power of the commun...

Страница 27: ......

Страница 28: ...he customer support for consultation 5 When performing warranty service if necessary we can request quali ed technicians identi ed by us to provide product circuit diagrams and repairable component da...

Отзывы: