02
1. Open flames or smoking during oxygen therapy are dangerous and may result in fire, facial burns, or
death. Do not allow smoking or open flames within 2m of the device or any oxygen carrying accesso-
ries, such as an attached cannula.
2. Do not use oil, grease or petroleum-based products on or near the device as the use of such products
may damage the electronic components of the device.
3. For proper operation, the device requires unobstructed ventilation.The device should be located as
to avoid pollutants or flames.
4. Do not plug or unplug the power cord when there is water on your hand. Otherwise it may cause
electric shock.
5. Do not use alcohol, solvents, polishes or any oily substances on the device. Otherwise, the output
oxygen may be contaminated, causing injury or physical discomfort.
6. When placing the oxygen concentrator, please keep a distance of 16cm from walls, curtains or other
objects, otherwise it may affect the air exchange of the air exhaustion of oxygen concentrator.
7. The device should not be used adjacent to or stacked with other equipment because it could result in
improper operation .
8. Do not use the oxygen concentrator if either the plug or power cord is damaged . Do not use
extension cords or electrical adapters.
9. Do not attempt to clean the concentrator while it is plugged into an electrical outlet.
10. The oxygen concentrator regularly emits a "ta..pu" exhaust sound when the oxygen concentrator
is operating (about 7 seconds).
11. The manufacturer recommends that the oxygen concentrator runs for no less than 30 minutes
each time when turned on. Do not turn on and off the oxygen concentrator frequently. It is suggested
to turn on/ turn off after 3 to 5 minutes each time, to avoid affecting the life of the compressor.
12. The use of accessories or service equipment not specified for use with the device may impair its
performance.
13. Please do not adjust the flow rate above 5L/min, otherwise the oxygen purity may not reach more
than 90%.
14. When the oxygen purity is abnormal, the user should contact the customer support for consulta-
tion (see the back cover).
15. If you begin to feel ill or are experiencing discomfort while using the oxygen concentrator, please
go to the doctor immediately.
16. When the user is a child or a person with limited mobility, the device needs to be used under the
supervision of the custodian.
17. If you have plan not to use the oxygen concentrator for several days, please clean and dry the
humidification bottle for storage.
18. Molecular sieve is a consumable, it will absorb moisture from the air during use and storage, and
its performance will decay, resulting in a decrease in oxygen purity. Before the first use, if the oxygen
purity is lower than 90%, please contact the customer support for consultation (see the back cover).
19. Do not leave the nasal oxygen cannula on bed covering or chair cushion, if the oxygen concentra-
tor is turned on but not in use, the oxygen will make the materials flammable. Turn the device off
when it is not in use.
20. The meaning of symbols.
Stacking layer limit
Manufacturer
This way up
Alternating current
No smoking
Keep dry
Class Ⅱ equipment
Consult instructions
for use
Symbol Meaning
Symbol Meaning
Symbol Meaning
Indicates a warning
or note
Temperature limit
Humidity limitation
Atmospheric
pressure limitation