Ⅴ. PRODUCT INTRODUCTION
The Y-506W oxygen concentrator uses the principle of Pressure Swing Adsorption to separate the
oxygen in the air from nitrogen and other gases. At room temperature, the oxygen which meets
medical standards can be continuously separated from the air when the power is turned on.
Product composition: Main unit, Power cord, Instruction manual, Nasal oxygen cannula, Filter
sponge, Disposable kit for nebulization, Fuse, Humidification bottle, Humidification tube,
Screwdriver, Warranty card, Instructions for unpacking.
Ⅶ. TRANSPORTATION AND STORAGE CONDITIONS
1. Ambient temperature range: -20℃~+55℃
2.Relative humidity range: ≤93%, and no condensation
3. Atmospheric pressure range: 700hPa~1060hPa
Note: The oxygen concentrator should be stored in an environment like this: No strong sunlight,
no corrosive odor, well ventilated, and clean. Avoid violent vibration, inverted or horizontal
placement during transportation.
Ⅷ. THE SCOPE OF APPLICATION OF THE PRODUCT
Using air as the raw material, the molecular sieve pressure adsorption process is used to produce
oxygen (Oxygen purity 90% to 96%). Could be used for oxygen therapy or liquid nebulization
treatment.
Ⅵ. PRODUCT USE ENVIRONMENT AND NORMAL
WORKING CONDITIONS
1. Power requirements: AC220V ± 22V, 50Hz ± 1Hz
2. Ambient temperature: 5℃~40℃
3. Relative humidity: ≤80%
4. Atmospheric pressure: 860hPa~1060hPa
5. Warm-up time: 30min
Note: If the storage or transportation temperature was lower than 5°C, it is recommended to place
the device in a normal working condition for more than 4 hours before using.
IPX0
Type BF applied part
No open flames
Symbol Meaning
Symbol Meaning
Symbol Meaning
On (main power)
No protection
Off (main power)
Fragile items