Setting Measurement Items
ZFX-C User’s Manual
3
SETTING THE MEAS
UREM
ENT CONDITIONS
91
Possible output results
The following values can be output when expressions are set.
Nos. indicating hue, saturation and value p.230
Item
Description
Message
Judgment result
The judgment result is output. (0:OK, -1:NG, -2: not measured)
Judge (JG)
Hue
The hue average is output. (0 to 360.0)
H average (H)
Saturation
The saturation average is output. (0 to 100.0)
S average (S)
Value
The value average is output. (0 to 100.0)
V average (V)
Hue deviation
The hue fluctuation is output. (0 to 180.0)
H deviation (HD)
Saturation deviation
The saturation fluctuation is output. (0 to 50.0)
S deviation (SD)
Value deviation
The value fluctuation is output. (0 to 50.0)
V deviation (VD)
Reference hue
The hue when the measurement region is set is output.
(0 to 360.0)
Ref. H average (SH)
Reference saturation
The saturation when the measurement region is set is output.
(0 to 100.0)
Ref. S average (SS)
Reference value
The value when the measurement region is set is output.
(0 to 100.0)
Ref. V average (SV)
Hue difference
The hue difference obtained by "measurement value - reference
value" is output. (-360.0 to 360.0)
H average dif. (DH)
Saturation difference
The saturation difference obtained by "measurement value -
reference value" is output. (-100.0 to 100.0)
S average dif. (DS)
Value difference
The value difference obtained by "measurement value - reference
value" is output. (-100.0 to 100.0)
V average dif. (DV)
Reference hue deviation The hue fluctuation when the measurement region is set is
output. (0 to 180.0)
Ref. H deviation (SHD)
Reference saturation
deviation
The saturation fluctuation when the measurement region is set is
output. (0 to 50.0)
Ref. S deviation (SSD)
Reference value
deviation
The value fluctuation when the measurement region is set is
output. (0 to 50.0)
Ref. V deviation (SVD)
Hue deviation difference The hue deviation difference obtained by "measurement value -
reference value" is output. (-180.0 to 180.0)
H deviation dif. (DH)
Saturation deviation
difference
The saturation deviation difference obtained by "measurement
value - reference value" is output. (-50.0 to 50.0)
S deviation dif. (DSD)
Value deviation
difference
The value deviation difference obtained by "measurement value -
reference value" is output. (-50.0 to 50.0)
V deviation dif. (DVD)
Hue maximum/minimum The hue maximum/minimum values are output. (0 to 360.0)
Max. H/Min. H
(HH, LH)
Saturation maximum/
minimum
The saturation maximum/minimum values are output. (0 to 100.0) Max. S/Min. S
(HS, LS)
Value maximum/mini-
mum
The value maximum/minimum values are output. (0 to 100.0)
Max. V/Min. V
(HV, LV)
Содержание Smart Sensor ZFX-C
Страница 2: ......
Страница 125: ...Additional Functions 122 ZFX C User s Manual MEMO ...
Страница 135: ...Checking Adjusting the Measurement ADJ Mode 132 ZFX C User s Manual MEMO ...
Страница 221: ...Menu List 218 ZFX C User s Manual MEMO ...
Страница 225: ...INDEX 222 ZFX C User s Manual MEMO ...
Страница 226: ...INDEX ZFX C User s Manual 8 APPENDICES 223 MEMO ...
Страница 227: ...INDEX 224 ZFX C User s Manual MEMO ...