5.Handling Errors and Problems
5. Handling Errors and Problems
5.1
Error Messages
Error Display
Cause
Remedy
Cuff is under inflated.
Press air release button and
restart the measurement with a
higher inflation level. Carefully
read steps under section 3.3.
Movement during
measurement
Repeat measurement. Remain
still and do not talk during
measurement.
Refer to section 3.3.
Air tube disconnected.
Insert the air tube securely.
Refer to section 3.2.
Arm cuff not applied
correctly.
Apply the arm cuff correctly.
Refer to section 3.2.
Clothing is interfering
with the arm cuff.
Remove any clothing
interfering with the arm cuff.
Refer to section 3.2.
Air is leaking from the
arm cuff.
Replace cuff with new one.
Refer to Chapter 7.
The arm cuff was
inflated above
299 mmHg.
Release your hand from the air
inflation bulb before the
pressure reaches 299 mmHg.
Refer to section 3.3.
Battery power is low.
Replace all four “AAA”
batteries with new ones.
Refer to section 2.1.
Blinks or
appears
continuously
Care and maintenance