background image

K3TJ

K3TJ

9

2. Display Shift Adjustment

The display can be shifted within a range of --1,999 to 9,999.
Adjust the display shift while applying an actual input.

Display after shifting

4 mA or 1 V

4- to 20-mA or 1- to 5-V range

Display after shifting

Input

Input

0- to 5-V or 0- to 10-V range

1000

1000 + b

1000 + b

1000

b

0

b

0

Example: When shifting from

10

to

0

Use the Up or Down Key to change the display. To
change the value to

0

, use the Down Key. The displayed

value will increase automatically if the Up Key is pressed
for 1 s or more and will decrease automatically if the
Down Key is pressed for 1 s or more.

Up Key or
Down Key

10

1

3
4
5
6
7
8

2

OFF

ON

0

Set parameter selector pin 2 to ON, at which time the
display will not change but a display shift adjustment will
be possible.

After

0

is displayed, set parameter selector pin 2 to OFF

to complete the display shift adjustment.

The display will be shifted
by --10.

1

3
4
5
6
7
8

2

OFF

ON

Display

Shift adjustment

Input

0

10

Содержание K3TJ

Страница 1: ...htness can be adjusted Conforms to the EMC standard EN61010 1 IEC1010 1 uC Ordering Information DC Voltage Input Model Voltage measurement range Display color Supply voltage g g p y 100 to 240 VAC 24...

Страница 2: ...nce Malfunction 10 to 55 Hz 0 5 mm single amplitude for 10 minutes each in X Y and Z directions Destruction 10 to 55 Hz 0 75 mm single amplitude for 2 hours each in X Y and Z directions Shock resistan...

Страница 3: ...panel meets the requirements of IP50 instead Measuring Ranges Measuring range Input impedance Reliability Max permissible continuous load Voltage DC 1 to 5 VDC 1 M 0 1 rdg 1 digit 250 V g 0 to 5 VDC 1...

Страница 4: ...iver into the groove on the bottom of the front panel and pull the front panel off The following is seen after the front panel is removed Parameter selector Up Key Down Key 1 34 5 6 7 8 2 ON No Name P...

Страница 5: ...o 1000 0 to 10 VDC input DC Current Input Model Input range Display 4 to 20 mA DC input 0 to 1000 1000 0 max min Display Input Adjust the display shift and scaling for the K3TJ while applying an actua...

Страница 6: ...the display To change the value to 0 use the Down Key The displayed value will increase automatically if the Up Key is pressed for 1 s or more and will decrease automatically if the Down Key is press...

Страница 7: ...er selector pin 3 to ON at which time the display will not change but a scaling adjustment will be possible After 1200 is displayed set parameter selector pin 3 to OFF to complete the scaling adjustme...

Страница 8: ...gntimes This methodis effectivewhentheuserrequiresa long display refresh period Inmovement averageprocessing method the samplingdata obtainedfrom samplingn times including thepresent samplingoperation...

Страница 9: ...mit setting range off No zero limit factory set Zero Limit Setting Range Display Zero limit setting value Input 7 Step Display Setting Step display setting is used to change the step of the displayed...

Страница 10: ...alue stored in the non volatile memory has been corrupted Execute display shift again according to the display shift adjustment procedure e 12 The scaling value stored in the non volatile memory has b...

Страница 11: ...ections 24 VDC HOLD GND GND 0 to 10 VDC 0 to 5 VDC 1 to 5 VDC 4 to 20 mA DC INPUT POWER 3 2 1 4 5 6 7 8 9 100 to 240 VAC Note Terminals 3 and 6 are connected internally The power supply terminals are...

Страница 12: ...le Never use the K3TJ in locations where corrosive gas particularly sulfuric or ammonia gas exists Avoid use of the K3TJ in locations subject to severe shock or vibra tion excessive dust or excessive...

Страница 13: ...n Phone 06 949 6012 Fax 06 949 6021 ALL DIMENSIONS SHOWN ARE IN MILLIMETERS To convert millimeters into inches multiply by 0 03937 To convert grams into ounces multiply by 0 03527 Cat No N080 E1 1C In...

Отзывы: