Appendix E
Setting Examples
107
With the K3NX-AA
j
C-L4, the input of 0 to 10 A can be converted and displayed within a range of 0.000 to 10.000
and can be output in the range of 4 to 20 mA in proportion to the change in display value according to the following
conditions.
Output Pattern Selection: Level output
LL output for a display value of larger than 4.000
L output for a display value of larger than 6.000
H output for a display value of larger than 8.000
HH output for a display value of larger than 10.000
Setting value menu
Scaling menu
Setup menu
Option menu
Input range
1 s
Set with:
Power ON
RUN mode
Set with:
Set with:
HH set value
H set value
L set value
LL set value
1 s
Average processing
Comparative output
pattern
L linear output range
H linear output range
Set with:
Set with:
Set with:
Set with:
Set with:
Set with:
1 s