How to Measure Blood Pressure
(in MANU. Mode)
1. Wrap the arm cuff over the
patient’s arm.
3. Push the ON/OFF (power)
Button to turn on the power.
ON/OFF
AVG./1st/2nd/3rd
DEFLATION
START
STOP
HIDE
P-SET
100
AUTO
140
180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE
5. Set the P-SET (pressure set-
ting) Volume to “AUTO” or the
target inflation value.
•
When setting the P-SET to
“AUTO”, turn the Volume coun-
terclockwise as far as it goes un-
til you can hear the click sound.
•
When the systolic pressure is
expected to exceed 220 mmHg,
proper inflation may not be pos
-
sible in “AUTO” setting.
Set the target inflation value to
30 to 40 mmHg higher than the
expected systolic pressure.
ON/OFF
AVG./1st/2nd/3rd
DEFLATION
START
STOP
HIDE
P-SET
100
AUTO
140
180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE
7. Push the ON/OFF (power)
Button to turn off the power.
ON/OFF
AVG./1st/2nd/3rd
DEFLATION
START
STOP
HIDE
P-SET
100
AUTO
140
180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE
2. Place the stethoscope on the
patient’s arm.
ON/OFF
AVG./1st/2nd/3rd
DEFLATION
START
STOP
HIDE
P-SET
100
AUTO
140
180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE
4. Set the MODE Selector to
“MANU.”.
ON/OFF
AVG./1st/2nd/3rd
DEFLATION
START
STOP
HIDE
P-SET
100
AUTO
140
180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE
6. Push the START Button to
start the measurement.
•
In the “AUTO” setting, the unit
stops inflation at the pres
-
sure considered to be 30 to
40 mmHg above the expected
systolic pressure, then start
deflation.
•
The unit can be re-inflated
only while the START Button is
pushed.
•
The pressure display is syn-
chronized with the pulse rate
and renewed.
•
Each time the DEFLATION
(deflation control) Button is
pushed, the unit deflates by
5~10 mmHg.
•
When you press the STOP But-
ton, the unit deflates fast and
ends the operation.
ON/OFF
AVG./1st/2nd/3rd
DEFLATION
START
STOP
HIDE
P-SET
100
AUTO
140
180
220
260
280
MODE
AVG.
MANU.
CHECK
SINGLE