6. Specifications
Product description
Automatic Upper Arm Blood Pressure Monitor
Product category
Electronic Sphygmomanometers
Model (code)
M2 Basic (HEM-7121J-E)
X2 Basic (HEM-7121J-EO)
Display
LCD digital display
Cuff pressure range
0 to 299 mmHg
Pulse measurement range 40 to 180 beats / min.
Blood pressure measurement range 20 to 280 mmHg
Accuracy
Pressure: ±3 mmHg / Pulse: ±5% of display reading
Inflation
Automatic by electric pump
Deflation
Automatic pressure release valve
Measurement method
Oscillometric method
Operating mode
Continuous operation
IP classification
Monitor: IP20 / Optional AC adapter: IP21 (HHP-CM01) or IP22 (HHP-BFH01)
Rating
DC6 V 4 W
Applied part
Type BF (arm cuff)
Power source
4 “AA” batteries 1.5 V or optional AC adapter
(INPUT AC 100 - 240 V 50 - 60 Hz 0.12 - 0.065 A)
Battery life
Approximately 1000 measurements (using new alkaline batteries)
Durable period (Service life)
Monitor: 5 years / Cuff: 5 years / Optional AC adapter: 5 years
Operating conditions
+10 to +40 °C / 15 to 90% RH (non-condensing) / 800 to 1060 hPa
Storage / Transport conditions
-20 to +60 °C / 10 to 90% RH (non-condensing)
Contents
Monitor, arm cuff (HEM-CR24), 4 “AA” batteries, Instruction Manual and
Protection against electric shock
Internally powered ME equipment (when using only batteries)
Class II ME equipment (optional AC adapter)
Weight
Monitor: approximately 250 g (not including batteries) / Arm cuff: approximately 130 g
Dimensions (approximately value) Monitor: 112 mm (W) × 82 mm (H) × 140 mm (L)
Arm cuff: 145 mm × 466 mm (air tube: 600 mm)
Cuff circumference applicable to
the monitor
170 to 420 mm (included arm cuff: (220 to 320 mm))
Memory
Last measurement
Note
• These specifications are subject to change without notice.
• This monitor is clinically investigated according to the requirements of ISO 81060-2:2013 (excluding pregnant and pre-eclampsia patients). In the
clinical validation study, K5 was used on 85 subjects for determination of diastolic blood pressure.
• IP classification is degrees of protection provided by enclosures in accordance with IEC 60529. This monitor and optional AC adapter are
protected against solid foreign objects of 12.5 mm diameter and greater such as a finger. The optional AC adapter HHP-CM01 is protected
against vertically falling water drops which may cause issues during a normal operation. The optional AC adapter HHP-BFH01 is protected
against oblique falling water drops which may cause issues during a normal operation.
• Operating mode is classification in accordance with IEC 60601-1.