Characteristics
Setting accuracy
(
+
0.3% of set value or
+
1
%
C, whichever greater),
+
1 digit max.
Indication accuracy
(
+
0.3% of set value or
+
1
%
C, whichever greater),
+
1 digit max. (Set value coincides with the
indicated value, because no relative error exists between both values.)
Disable band
0.1% to 10.0% (in units of 0.1%)
Proportional band
0.1
%
to 999.9
%
C/
%
F (in units of 0.1
%
C/
%
F)
Integral time
1 to 3,999 s (in units of 1 s)
Derivative time
0 to 3,999 s (in units of 1 s)
Alarm output setting range
Thermocouple: –999
%
to 9,999
%
C/
%
F (in units of 1
%
C/
%
F)
Platinum resistance thermometer: –99.9
%
to 999.9
%
C/
%
F (in units of 0.1
%
C/
%
F)
Alarm output hysteresis
0.0
%
to 999.9
%
C/
%
F (in units of 0.1
%
C/
%
F)
Sampling period
500 ms (temperature sensor input)
Output refresh period
500 ms
Display refresh period
500 ms (temperature sensor input)
Valve travel indication
–99% to 999% (in units of 1%)
Full Open/close time
1 to 999 s ( in units of 1 s)
Input shift
Thermocouple: –999
%
to 9,999
%
C/
%
F (in units of 1
%
C/
%
F)
Platinum resistance thermometer: –99.9
%
to 999.9
%
C/
%
F (in units of 0.1
%
C/
%
F)
Insulation resistance
20 M
W
min. (at 500 VDC)
Dielectric strength
2,000 VAC 50/60 Hz for 1 min between terminals of different polarities
Vibration resistance
Malfunction: 2 to 55 Hz, 2G 10 min each in X, Y, and Z directions
Destruction: 10 to 55 Hz, 0.75-mm double amplitude 2 hrs each in X, Y, and Z directions
Shock resistance
Malfunction: 200 m/s
2
3 times each in 6 directions
Destruction: 300 m/s
2
3 times each in 6 directions
Life expectancy
Mechanical: 10,000,000 operations min.
Electrical:
100,000 operations min.
Ambient temperature
Operating: –10
%
to 55
%
C (with no icing)
Storage:
–25
%
to 65
%
C (with no icing)
Ambient humidity
Operating: 35% to 85%
Memory protection
Non-volatile memory
Enclosure ratings
Front panel: IEC standard IP50
Rear case: IEC standard IP20
Terminals:
IEC standard IP00
Weight
Standard: Approx. 350 g; Communications: Approx. 480 g; Mounting bracket: Approx. 50 g
*The accuracy of U thermocouples under temperatures from –150
%
to 400
%
C (–240
%
to 700
%
F) is
+
2
%
C (
+
3.6
%
F)
+
1 digit. Accuracy is
reduced below –150
%
C (–240
%
F). The accuracy of R and S thermocouples under 0
%
to 200
%
C (0
%
to 400
%
F) is
+
3
%
C (
+
5.4
%
F)
+