Characteristics
Setting accuracy
+
0.3% FS
+
1 digit max.
Indication accuracy
+
0.3% FS
+
1 digit max. (Set value coincides with the indicated value, because no relative
error exists between both values.)
Indication range
–10% to 110% FS
Hysteresis
0.0% to 100.0% FS (in units of 0.1% FS)
Proportional band
0.0% to 999.9% FS (in units of 0.1% FS)
Integral time
0 to 3,999 s (in units of 1 s)
Derivative time
0 to 3,999 s (in units of 1 s)
Alarm output setting range
–999 to 9999 (Decimal point is displayed at set position)
Alarm output hysteresis
0.0% to 100.0% FS (in units of 0.1% FS)
Scaling setting range
–999 to 9999 (Decimal point is displayed at set position)
Manual output setting range
0% to 100% (in units of 1%)
Control period
Pulse output: 1 to 99 s (in units of 1 s)
Sampling period
500 ms
Output refresh period
Pulse output:
500 ms
Current output: 500 ms
Display refresh period
500 ms
Shift set
–999 to 9999 (Decimal point is displayed at set position)
Input shift
–999 to 9999 (Decimal point is displayed at set position)
Insulation resistance*
20 M
W
min. (at 500 VDC)
Dielectric strength*
2,000 VAC 50/60 Hz for 1 min between terminals of different polarity
Vibration resistance
Malfunction: 2 to 55 Hz, 2G 10 min each in X, Y, and Z directions
Destruction: 10 to 55 Hz, 0.75-mm double amplitude 2 hrs each in X, Y, and Z directions
Shock resistance
Malfunction: 200 m/s
2
3 times each in 6 directions
Destruction: 300 m/s
2
3 times each in 6 directions
Ambient temperature
Operating: –10
%
to 55
%
C (with no icing)
Storage:
–25
%
to 65
%
C (with no icing)
Ambient humidity
Operating: 35% to 85%
Memory protection
Non-volatile memory
Enclosure ratings
Front panel: IEC standard IP50
Rear case: IEC standard IP20
Terminals:
IEC standard IP00
Weight
Standard: Approx. 350 g; Communications: Approx. 480 g; Mounting bracket: Approx. 50 g
*Tested with an Output Unit mounted.
Output Unit Characteristics
Relay unit life expectancy
Mechanical: 10,000,000 operations min.
Electrical:
100,000 operations min.