manualshive.com logo in svg
background image

CSM_61F-GP-N

@

_DS_E_7_2

1

Floatless Level Switch (Compact, Plug-in Type)

61F-GP-N

@

Space-saving Design Ideal for Control Panel 

Downsizing. Easy Maintenance.

• Compact: 49.4 

×

 38 

×

 84 mm (H

×

W

×

D).

• Easy identification of operating status with LED operation indicator.
• Independent DPDT contacts on 11-Pin Models.
• CE marking and UL/CSA compliance.

Model Number Legend

Ordering Information

Note:

When ordering, specify the desired operating voltage at the end of the model number.

1. No. of Pins

2. Type

N: 11 pins
N8: 8 pins

Blank: General-purpose
L 2KM: Long-distance (for 2 km)
L 4KM: Long-distance (for 4 km)
H:

High-sensitivity

D:

Low-sensitivity

R:

Two-wire

T: High-temperature

Refer to 

Safety Precautions for Floatless Level 

Controllers

.

Position of LED indicator

2

1

61F-GP-

@@

Type

General-purpose

Long-distance (for 2 km)

Long-distance (for 4 km)

Model

Model

Model

11-pin

61F-GP-N

61F-GP-NL 2KM

61F-GP-NL 4KM

Type

High-sensitivity

Low-sensitivity

Two-wire

Model

Model

Model

11-pin

61F-GP-NH

61F-GP-ND

61F-GP-NR

Type

Tropical environments

High-temperature

Model

Model

8-pin

61F-GP-N-TDL

61F-GP-NT

Type

General-purpose

Long-distance (for 2 km)

Long-distance (for 4 km)

Model

Model

Model

8-pin

61F-GP-N8

61F-GP-N8L 2KM

61F-GP-N8L 4KM

Type

High-sensitivity

Low-sensitivity

Two-wire

Model

Model

Model

8-pin

61F-GP-N8H

61F-GP-N8D

61F-GP-N8R

61F-GP-N8HY

Example: 61F-GP-N [220 VAC]

Desired supply voltage

Содержание 61F-GP-N8 -

Страница 1: ...High sensitivity D Low sensitivity R Two wire T High temperature Refer to Safety Precautions for Floatless Level Controllers Position of LED indicator 2 1 61F GP Type General purpose Long distance fo...

Страница 2: ...rature is high For control of ordi nary purified water in cases where the distance between sewage pumps and water tanks or between receiver tanks and supply tanks is long or where remote con trol is r...

Страница 3: ...1F GP N NT NL ND 61F GP NR 61F GP NH 3 9 1 4 10 11 U 24 V 8 V Control circuit S0 S1 E3 5 E1 Ta1 Tc1 Tb1 8 6 7 U Ta2 Tc2 Tb2 3 9 1 4 10 11 U 24 V 8 V Control circuit S0 S1 E3 5 E1 Ta1 Tc1 Tb1 8 6 7 U T...

Страница 4: ...mmercial Voltage 61F GP N PS 3S 3 9 E2 E3 E1 P 8 V 0 V U U U 24 V Control circuit See note 4 8 7 6 5 10 11 1 2 Start Stop Contac tor Power supply Motor protection relay Waste water tank Note Be sure t...

Страница 5: ...U U 61F GP N 5 6 7 8 4 8 V 0 V 24 V 10 3 9 1 2 11 Control circuit U See note Water tank Note The power supply phases terminals 3 to 9 can be matched to use the same ground for the common Electrode the...

Страница 6: ...Control circuit U2 B PL PL See note Upper limit Lower limit Alarm Water tank Motor protection relay Note The power supply phases terminals 3 to 9 can be matched to use the same ground for the common...

Страница 7: ...oir MCCB R S M T 61F GP NR See note 5 4 P R 8 V 0 V 24 V PS 3SR 10 3 9 1 2 11 Control circuit E3 Contac tor T Commercial Voltage Power supply Motor protection relay Waste water tank Note Be sure to gr...

Страница 8: ...20 or 240 VAC 61F GP N 220 or 240 VAC Note 1 The diagram shows the connections for the water supply When draining change the connection from terminal 1 to terminal 11 2 Be sure to ground terminal 4 Po...

Страница 9: ...r control of liq uids with high spe cific resistance such as distilled water For control of liq uids with low spe cific resistance such as salt water sewage water acid chemicals alkali chemicals For c...

Страница 10: ...ply Control circuit 24 V Power supply 8 V see note 61F GP N8 N8L N8D N8HY 61F GP N8H 61F GP N8R 24 V 8 V Power supply Control circuit 2 2 4 3 61F GP N8Y 24 V 8 V Control circuit Power supply Note 24 V...

Страница 11: ...N8 P 8 V 0 V U U U 24 V See note 3 4 5 6 PS 3S E2 E3 E1 Start Stop 2 1 8 7 Control circuit Contac tor Power supply Motor protection relay Waste water tank Commercial Voltage Note Be sure to ground the...

Страница 12: ...T 61F GP N8R See note 7 8 1 2 P R 8 V 0 V 24 V PS 3SR 3 5 6 4 Control circuit Stop Start E1 Contac tor Commercial Voltage Power supply Motor protection relay Waste water tank Note Be sure to ground t...

Страница 13: ...r to the following diagrams Line voltage R S S T or R T 380 or 415 VAC Phase voltage N R N S or N T 220 or 240 VAC 61F GP N8 220 or 240 VAC Note Be sure to ground terminal 1 Power source 380 to 415 VA...

Страница 14: ...t provided with the Level Controller t 0 3 4 5 83 4 6 Approx 80 6 2 8 5 9 32 5 9 0 D 216 8037 I21 2 7 66 9 6 I7 FREQ 50 60 Hz SECON DARY 8 V AC VOLTA E 38 70 84 49 4 61F GP N8 PF083A 91 61F GP N8 N8L...

Страница 15: ...ther in no event shall liability of Omron Companies exceed the individual price of the Product on which liability is asserted Suitability of Use Omron Companies shall not be responsible for conformity...