background image

16

Omni Tip™ Adapter

(A1000SB)

(Black Nose Ring)

5.4 AUTOCLAVING OMNI TIPS™

Omni Tips™ generator probes may be autoclaved up to seven times 

and retain their mechanical properties. Due to different combinations 

of chemicals and environments, it is suggested that the autoclaving 

recommendations be followed as a guideline only, and the user should 

decide how to best alter the procedure for their specific application.
The tube and shaft of Omni Tips™ can be autoclaved at up to a   

maximum of 250°F for 45 minutes. This includes a 15-minute ramp up, 

15-minute hold, and a 15-minute ramp down of the temperature cycle. 

If the tube or shaft begins to warp as a result of autoclaving, discard 

and lower the cycle time or temperature level accordingly for future 

runs.

5.5 CLEANING THE OMNI TIP™ ADAPTER

The Omni Tip™ (12-120 and A1000SB) adapters are constructed of 

aluminum with stainless steel rotating components. For general 

cleaning, it is recommended that the adapter be wiped down 

with a lint free cloth and an alcohol solution. Adequate molecular 

contamination cleaning can also be accomplished by scrubbing with 

soap and hot water, followed by brief rinsing with 1% SDS, and then 

by distilled water which has been treated with diethylpyrocarbonate 

(DEPC). 

Autoclaving, baking, or submersion in cleaning agents is 

not recommended;

 this will shorten the life of the adapter housing 

and bearings.

SECTION 5 — OMNI TIP™ PLASTIC PROBES

Omni Tip™ Adapter

(12-120)

(Blue Nose Ring)

Содержание PCR-258

Страница 1: ...REV B PCR 258 OMNI Tip HOMOGENIZING KITS USER MANUAL...

Страница 2: ...ithout charge For the balance of the period parts will be provided but service will be charged at established labor rates This warranty does not include normal wear from use it does not apply to any i...

Страница 3: ...ion fire or injury Use this product only for its intended purpose as described in this booklet DO NOT use attachments not recommended by the manufacturer DO NOT operate the product if it is damaged in...

Страница 4: ...for those applications WARNING Keep all housings in place and in working order WARNING Remove all tools from the generator probe before turning the motor on WARNING DO NOT use the motor in a dangerou...

Страница 5: ...he Omni TH 3 1 Stand Mounting the Omni TH 9 3 2 Installing Rotor Stator Generator Probes 9 3 3 Omni Tip Plastic Probes 10 3 4 Omni TH Operation 10 3 5 Theory of Operation 11 Section 4 Operational Guid...

Страница 6: ...n for more information Omni s Tissue Homogenizing Kits are designed for situations where molecular contamination between samples cannot be tolerated Stainless steel generator probes are available for...

Страница 7: ...ssue Homogenizing Kits is required for mounting Omni Tips to the Omni TH Homogenizer and is not disposable See Section 5 for more information about Omni Tips Part Number A1000SB 12 120 30750 30750H 32...

Страница 8: ...03 100 mL Height motor only 22 9 cm 9 in Weight motor only 466 g 16 4 oz Electrical Requirements 115 V 60 Hz or 220 V 50 Hz Standards Approval Compliance 220V CE Certified Description Motor Drive Unit...

Страница 9: ...n firmly The unit will lock in place 4 To remove the TH press the black button on the back of the opening of the post clamp assembly and lift the motor from the cradle 3 2 INSTALLING ROTOR STATOR GENE...

Страница 10: ...he sample vessel can significantly affect flow patterns within the vessel This also affects processing efficiency As a rule of thumb the generator probe usually operates most efficiently at a depth of...

Страница 11: ...eraction of the rotor knife with these windows sets up a shearing action reducing the particle size of the sample The speed differential between the rapidly moving portion and the relatively stationar...

Страница 12: ...n times for repeated use they are not recommended for extended use with phenol or guanidine based reagents see Section 5 3 for chemi cal resistance compatibility Disposal may be required when used in...

Страница 13: ...s Processing Speed Omni Tip Homogenizing Kits include a handheld Omni TH Homogenizer with motor speeds up to 35 000rpm Operating at the top speed enhances processing efficiency of the generator probes...

Страница 14: ...ntamination for many applica tions the probe can be cleaned sterilized or disinfected and reused by the procedures described below NOTE Omni Tip plastic probes are not factory presterilized or disinfe...

Страница 15: ...euse However phenol or guanidine based reagents commonly used in PCR studies are not recommended for extended use with Omni Tips Exposure to these chemicals shortens the life of Omni Tips and usually...

Страница 16: ...ft begins to warp as a result of autoclaving discard and lower the cycle time or temperature level accordingly for future runs 5 5 CLEANING THE OMNI TIP ADAPTER The Omni Tip 12 120 and A1000SB adapter...

Страница 17: ...he Omni TH is plugged in and turned on but is not functioning Check that the brushes are not worn and are correctly installed Replace if necessary see Section 3 2 Motor is turned ON and makes a makes...

Страница 18: ...18 NOTES...

Страница 19: ...19 NOTES...

Страница 20: ...www omni inc com OMNI Tip HOMOGENIZING KITS Omni International Inc 935C Cobb Place Blvd Kennesaw GA 30144 800 776 4431 770 421 0058...

Отзывы: