Installing a Battery
WARNING: replace the battery with a CR2430 coin cell only.
Remove battery if the device is not to be used for an extended
period of time, such as 1 month.
1. Turn the stimulator off and then back on.
2. Remove OML LINQ
TM
from the Holder if attached (see page 7).
3. Slide open the battery cover. The cover slides away from the
OML LINQ
TM
. Press down on the rear portion of the cover to
release the clip (please refer to image A and B).
4. If present then remove the old battery.
5. Install the new battery with the positive side of the battery
facing out from the OML LINQ
TM
. If installed correctly the blue
light will flash once (please refer to image C).
6. Slide the battery cover back on to the main body. The battery
cover is fully inserted when there are no gaps between the
casing and the battery cover (please refer to image D). Do not
use excessive force.
7. Re-insert the OML LINQ
TM
into the Holder (see page 6), insert
the Wired Footswitch and check that the stimulator unpauses
and responds to a press of the footswitch (see page 12).