23
Electromagnetic Compliance
Manufacturer’s declaration of electromagnetic immunity
The OML LINQ™ is intended for use in the electromagnetic
environment specified below. The customer or user of the OML
LINQ™ should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Com-
pliance
level
Electromagnetic
environment -
guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2
± 8kV
contact
± 15kV air
± 8kV
contact
± 15kV
air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Electrical fast
transient/burst
IEC 6100-4-4
No
connection
to mains
ODFS Pace® (XL)
does not have a
mains power supply
Surge
IEC 6100-4-5
No
connection
to mains
Voltage
dips, short
interruptions
and variations
IEC 61000-4-11
No
connection
to mains
Power
frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m Power frequency
magnetic
fields
should be at levels
characteristic of a
typical location in a
typical commercial,
hospital or domestic
environment.