High-Voltage Output
Conditions:
Signals below 45Hz with reduced values possible. Capacitive
linear loads.
11.1.2
Measurements
Test frequencies
TanDelta test card: Column "Hz" of the results table
Special displays in the frequency column "Hz" and their meanings:
Filter for selective measurements
Conditions:
f0 = 15 ... 400Hz
Terminal
U / f
THD
I
S
t
max
High volt.
output
10 ... 12kV AC
15 ... 400Hz
< 2%
300mA
3600VA
> 2min
100mA
1200VA
> 60min
Range
Resolution
Typical accuracy
15 ... 400Hz
0.01Hz
error < 0.005% of reading
*50Hz (*60Hz)
Measurement mode suppressing the mains frequency
interferences; doubles the measurement time.
!30Hz
The selected test voltage is not available in Automatic
measurement (applies to frequencies below 45Hz only).
?xxHz
Results with reduced accuracy, e.g., in case of a low testing
voltage, influences of partial discharge etc.
Filter Bandwidth Meas. time Stop band specification (attenuation)
f0 ± 5Hz
2.2s
> 110dB at fx = f0 ± (5Hz or more)
Содержание CP TD1
Страница 1: ...CP TD1 Reference Manual...
Страница 14: ...CP TD1 Reference Manual V 1 44 14...
Страница 20: ...CP TD1 Reference Manual V 1 44 20...
Страница 32: ...CP TD1 Reference Manual V 1 44 32...
Страница 38: ...CP TD1 Reference Manual V 1 44 38...
Страница 84: ...CP TD1 Reference Manual V 1 44 84 Figure 6 14 Frequency scan for H L f 5 kV...
Страница 114: ...CP TD1 Reference Manual V 1 44 114...
Страница 128: ...CP TD1 Reference Manual V 1 44 128...
Страница 154: ...CP TD1 Reference Manual V 1 44 154 Figure 12 12 Diverter switch for 110 kV 40 MVA...
Страница 184: ...CP TD1 Reference Manual V 1 44 184...
Страница 186: ...CP TD1 Reference Manual V 1 44 186...