4.1 Connecting the CMGPS to a CMC Test Set
Figure 4-1:
CMGPS
communication
interface
ext. Interf.
Figure 4-2:
Power supply unit
100 - 240 VAC/18 VDC
incl. plug-in adapter for
country-specific wall
outlets
1. Section 3.2 ”LED Status Display” on page 17 describes the state of the LED
signalization during the activation and initialization of the
CMGPS
.
The following explanations are based on successful initialization and on the
fact that the
CMGPS
locked itself to the GPS signal.
The state of the LEDs of the
CMGPS
should then be as follows:
1. Connect the
ext. Interf.
of the
CMGPS
to
the
ext. Interf.
input at the rear of the CMC
test set by means of the supplied
connection cable VEHK0003 (shown in
figure 3-3).
2. Plug in the DC adapter of
the 100-240 V
AC
/18 V
DC
power supply unit to the
DC in
power supply input of
the
CMGPS
.
3. Connect the power supply
unit to a wall outlet.
Note:
The power supply
unit is not needed with up-
to-date test sets such as
CMC 256
,
CMC 256plus
,
CMC 353, CMC 356
and
CMC 850.
In conjunction
with the
CMIRIG-B
interface
unit, use the above
mentioned power supply
unit to power the
CMGPS
.
O n
Pulse
Searching
Locked
LED "ON" is lit.
LED "Searching/Locked" is lit.
Содержание CMGPS
Страница 1: ...27 CMGPS Synchronization Unit Reference Manual...
Страница 8: ...GPS Synchronization Unit CMGPS 8...
Страница 10: ...GPS Synchronization Unit CMGPS 10...
Страница 24: ...GPS Synchronization Unit CMGPS 24...
Страница 28: ...GPS Synchronization Unit CMGPS 28...
Страница 34: ...GPS Syncronization Unit CMGPS 34...
Страница 46: ...Contact Information Technical Support 46...