Technical Data EP Option
49
6.1.1
Current Outputs CURRENT OUTPUT 1-3
3 current outputs
1
range 12.5A
Output currents
3-phase AC (L-N)
1-phase AC (L-N)
in
parallel
1-phase DC (L-N)
3
×
0 to 12,5 A
1
×
0 to 12, 5A
1
×
0 to 27 A at >45 Hz
1
×
0 to 21 A at <45 Hz
1
×
0 to
±
12,5 A
Power
2
3-phase AC (L-N)
1-phase AC (L-L)
3
1-phase AC (L-N)
4
1-phase DC (L-N)
typ. 3
×
43 VA
typ. 1
×
86 VA
typ. 1
×
65 VA
typ. 1
×
62 W
guar. 3
×
40 VA
guar. 1
×
75 VA
guar. 1
×
74 VA
guar. 1
×
59 W
Resolution
< 500 µA
Accuracy
5
typical error
< 0.015 % rd.
+ 0.005 % of rg.
guaranteed error
< 0.04 % rd.
+ 0.01 % of rg.
Harmonic distortion
6
typ. 0.025 %
guar. < 0.05 %
Short-circuit protection
Unlimited against N and from L to L.
Open-circuit protection
Open outputs (idling) allowed.
Connection
4 mm banana sockets or amplifier
combination socket.
Isolation
Reinforced isolation to mains and to all SELV
interfaces.
Table 30
Current outputs, range 12.5 A (EP option)
1
Data for 3-phase systems are valid under symm. conditions (0 °, 120 °, 240 °).
2
For higher current or output power demands connection of an external amplifier is possible (e.g.,
CMA current amplifier). The power data refers to an output current of 12.5 A
eff
or 12.5 A
DC
.
3
Single-phase model (operation with two phases in phase opposition):
two currents in series, additional adapter SPA1 or SPA156 recommended.
4
3 phases (3 x 9 A) in parallel for in-phase operation.
5
rd. = reading; rg. = range, whereat
n
% of rg. means:
n
% of upper range value (12.5 A).
6
Values at 50/60 Hz, 20 kHz measuring band width, nominal value and nominal load.