Test Set CMC 156
48
6.1 Outputs (EP option)
The power output specifications of test sets with EP option (Extended
Precision) may differ in substantial extent from the standard design. For the
standard design specifications, please refer to Paragraph 5.2 “Outputs”.
A block diagram of the generator outputs is given in Figure 3 on page 18.
Analog current and voltage outputs
Frequency range
1
Sinusoidal
signals
Transient
signals
10 to 1000 Hz
DC to 3.1 kHz
Frequency resolution
5 µHz
Frequency accuracy
0.5 ppm
Frequency drift
1 ppm
Phase range
- 360 ° to + 360 °
Phase resolution
0.001 °
Phase error
2
typ. 0.005 °
Guaranteed < 0.02 °
Table 29
Analog current and voltage outputs
1
If you purchased the option
FL-6
, the maximum output frequency is constrained to
599 Hz
.
2
Valid for sinusoidal signals with 50 / 60 Hz.