background image

 

 

OMICRON Lab  

13 

Technical Support 

 

Technical Support 

When you are working with our products we want to provide you with the greatest possible benefits. If 
you need any support, we are here to assist you! 

Technical Support - Get Support  

 

www.omicron-lab.com/support

 

[email protected]

 

 

 

At our technical support hotline, you can reach well-educated technicians for all your questions. 
Competent and free of charge. 

Make use of our technical support hotlines: 

Americas:   

 

+1 713 830-4660 or +1 800-OMICRON

 

Asia-Pacific: 

 

+852 3767 5500

 

Europe / Middle East / Africa: 

+43 59495 4444

 

Additionally, you can find the OMICRON Lab Service Center or Sales Partner closest to you at 

www.omicron-lab.com

 

→ Contact. 

 

 

 

 

 

 

 

 

 

 

 

 

OMICRON Lab  

OMICRON electronics GmbH,  

Oberes Ried 1, 6833 Klaus, Austria. +43 59495

Содержание B-WIC

Страница 1: ...uusohjeet Consignes de s curit Sigurnosne upute Biztons gi utas t sok Istruzioni di sicurezza Saugos nurodymai Dro bas instrukcijas Veiligheidsinstructies Instrukcje bezpiecze stwa Instru es de segura...

Страница 2: ...gnition and or storage in electronic data processing systems requires the explicit consent of OMICRON electronics Reprinting wholly or in part is not permitted The product information specifications a...

Страница 3: ...vi ena namjena i kvalifikacije rukovatelja 16 Biztons gi utas t sok rendeltet sszer haszn lat s kezel i szakk pes t si k vetelm nyek 16 Istruzioni di sicurezza utilizzo previsto e qualifiche degli ope...

Страница 4: ...annot be disposed of as household waste For customers in EU countries incl European Economic Area OMICRON test sets are subject to the EU Waste Electrical and Electronic Equipment Directive WEEE direc...

Страница 5: ...tages or currents 2 3 Designated Use B WIC and B SMC have been especially designed to measure the impedance of passive components in conjunction with the Bode 100 in laboratory or manufacturing enviro...

Страница 6: ...B SMC Fixture 6 pcs 100 0 1 Load Calibration Resistors Size 0603 Figure 3 1 B WIC and B SMC connector description SOURCE connector BNC input plug connect to the output of Bode 100 CH1 connector BNC ou...

Страница 7: ...ge signal proportional to the voltage at the DUT and CH2 receives a voltage signal proportional to the current flowing thru the DUT Figure 4 1 Measurement principle of B WIC and B SMC Open Short Load...

Страница 8: ...ingle sweep or a continuous sweep by clicking on the corresponding icons 4 4 Calibration Correction Before you can start a measurement Open Short Load calibration correction must be performed This mov...

Страница 9: ...calibration 4 4 2 Calibrating B SMC with a 100 resistor Open Short Load Use the knurled nut to adjust the distance of the electrode needle to the plate to match the DUT size to compensate for stray c...

Страница 10: ...ass the internal measurement bridge causing errors in the measurement The same is true for physically large DUTs Figure 4 4 Stray capacitance of leads and physically large DUTs 4 5 2 Electrode Dimensi...

Страница 11: ...setting Contact resistance Both the B WIC and B SMC are subject to contact resistance The gold plating of the electrodes lowers the contact resistance to around 0 5 m Make sure to have clean contacts...

Страница 12: ...MHz Recommended impedance measurement range 20 m to 600 k B CAL 3 load accuracy 100 0 1 15 ppm C 0603 load resistors 100 0 1 15 ppm C Table 5 2 Environmental conditions Characteristic Rating Operating...

Страница 13: ...com At our technical support hotline you can reach well educated technicians for all your questions Competent and free of charge Make use of our technical support hotlines Americas 1 713 830 4660 or 1...

Страница 14: ...f r die Messung der Impedanz von passiven Komponenten mit dem Bode 100 in Labor und Produktionsumgebungen vorgesehen Qualifikation des Bedienpersonals Pr fungen mit den Impedanzmessadaptern B WIC und...

Страница 15: ...C y B SMC est n especialmente dise ados para medir la impedancia de los componentes pasivos junto con el instrumento Bode 100 en entornos de laboratorio o de fabricaci n Cualificaciones de los operado...

Страница 16: ...avail avec l quipement Les essais effectu s l aide du B WIC ou du B SMC doivent tre conformes aux consignes de s curit internes et tout autre document pertinent Sigurnosne upute predvi ena namjena i k...

Страница 17: ...peratoriaus Bandymai su B WIC ir B SMCturi b ti atliekami laikantis vidaus saugos nurodym ir kit svarbi dokument Dro bas instrukcijas paredz t izmanto ana un operatora kvalifik cija Dro bas instrukcij...

Страница 18: ...ar com o equipamento O teste com o B WIC e o B SMC deve estar em conformidade com as instru es de seguran a internas e com os documentos relevantes adicionais Instruc iuni de siguran destina ia de uti...

Страница 19: ...ljavca Preizku anje z napravami B WIC in B SMC mora biti skladno z notranjimi varnostnimi navodili in dodatnimi ustreznimi dokumenti S kerhetsinstruktioner avsedd anv ndning och anv ndarkvalifikatione...

Страница 20: ...Version 1 0 info omicron lab com www omicron lab com OMICRON Lab 2020 B WIC B SMC User Manual English All rights reserved ENU 1248 05 01...

Отзывы: