Commissioning and operation
Series TCL-3M165E2
- 18 -
4.3.1 Dry block calibrator
Type
TCL-3M165E2
Dry block
Adapter sleeves with single or multiple holes are used for the calibration of straight
temperature sensors.
To achieve the specified accuracy of the calibrators (
§ 10.1) the temperature
sensor (test specimen) and the adapter sleeve must be matched to one another:
•
The borehole of the adapter sleeve may be a maximum of 0.5 mm (0.12 in)
larger than the diameter of the test specimen.
•
The measurement element of the test specimen must be located in the
homogeneous temperature zone of the adapter sleeve.
Inserting:
The appropriate adapter sleeve is inserted
into the metal block with the aid of the
sleeve exchange tool.
External reference sensor (optional):
Align the adapter sleeve so that the hole
for the external reference sensor is located
at 12 o’clock.
Removing and cleaning:
Let the calibrator cool before you remove
the adapter sleeve.
Pull the adapter sleeve out of the metal
block with the help of the sleeve exchange tool.
Clean the adapter sleeve and the metal block.
This prevents the adapter sleeve from getting stuck in the metal block.
Adapter sleeves and sleeve exchange tool