5.1.3.3
Lock
If set, the user specified units of measure string (4-character maximum) will be used in place of the default units of
measure.
5.1.3.4
Assigned
Refer to the
Smart Sensor Device Interface
documentation.
5.1.3.5
Available
Refer to the
Smart Sensor Device Interface
documentation.
5.1.4
Thermocouple Device Byte
The Sensor Device byte determines whether the Open Circuit detection is enabled.
I/O (TC)
7
6
5
4
3
2
1
0
0
0
0
Open Circuit Detect
0
Disabled
1
Enabled
0
0
0
0
5.1.5
IPSO Thermocouple Temperature Sensor Definition
The Thermocouple sensor IPSO definition provides signal range, measured min/max values, and IPSO object type
information. The Range information is Thermocouple Type dependent.
Offset
Name
Value
Description
0x00
Sensor Type
3303 Temperature (oC)
0x02
Precision
1
Provides reading of xxx.x
0x04
Sensor Trigger Function
??
See 4.1.5.1
0x08
Min Measured
??
Minimum reading since last reset
0x0c
Max Measured
??
Maximum reading since last reset
0x10
Min Range
??
Type
Min Range
Max Range
J
-210
1200
K
-160
1272
T
-190
400
E
-220
1000
N
-100
1300
R
40
1788
S
100
1768
B
640
1820
X
0
2320
0x14
Max Range
??
5.1.5.1
Sensor Trigger Function
The Sensor Trigger function is used to reset the IPSO min/max values as well as control the ICE Point / Single-Point Calibration
process
.
Sensor Trigger Function
7
6
5
4
3
2
1
0
0
0
0
0
0
0
0
Reset
Min/Max
15
14
13
12
11
10
9
8
0
0
Calibration
Reset
Calibration
Status
Calibration
Mode
0
Capture
Low
Calibration
Start