31
Configuring The Meter
4
4.11 CONFIGURING TEMPERATURE (C or F - “TEMP”)
If you have selected “
MANU
” in “
A.T.C
.” menu item (Section 4.1), then you have
to specify the constant temperature value. Maximum/minimum values are 0 to
199.9 for C or 0 to 398.0 for F.
1.
Press the
MENU
button until the meter shows “
TEMP
”.
2.
Press the
T/mV
button. The meter shows actual constant temperature value.
3.
Press the
/pH
button until the first digit is correct.
4.
Press the
T/mV
button to advance to the next digit.
5.
Repeat steps 3 and 4 until meter shows the desired “
TEMP
” value.
6.
Press the
MENU
button to store your selection. The meter momentarily shows
“
STRD
”, followed by “
OT.S.O
” (Output Scale and Offset) - refer to section 4.12.
7.
Press the
MENU
button until the meter shows the “
CAL.2
or
CAL.3
” menu.
If the display shows “
CAL.2
” then proceed to Section 4.4.
If the display shows “
CAL.3
” then proceed to Section 4.5.
C or F flashes to indicate the unit of measure.