Configuring The Meter
4
4.10 USING OUTPUT CONFIGURATION (OT.CF)
Use Output Configuration to select the following:
• To enable or disable the analog output
• To determine if the analog output is current or voltage
4.10.1 Enabling or Disabling the Analog Output
To enable or disable the analog output, follow these steps:
1.
Press the
MENU
button until the meter shows “
OT.CF
”.
2.
Press the
T/mV
button. The meter shows one of the following:
•
“O.1=E” (Analog output enabled)
(Default)
• “O.1=D” (Analog output disabled)
3.
Press the
/pH
button to toggle between available choices.
4.
Press the
T/mV
button to select the analog output as current/voltage or press the
MENU
button to store your selection. The meter momentarily shows “
STRD
”,
followed by “
OT.S.O
” (Output Scale and Offset) - refer to section 4.12.
4.10.2 Selecting Analog Output as Current or Voltage
1.
Press the
T/mV
button. The meter shows one of the following:
•
“O.2=C" (Analog output = current)
(Default)
• “O.2=V” (Analog output = voltage)
2.
Press the
/pH
button to toggle between available choices.
3.
Press the
T/mV
button to go back to selecting analog output as enabled or
disabled or press the
MENU
button to store your choices. The meter momentarily
shows “
STRD
”, followed by “
OT.S.O
” (Output Scale and Offset) - refer to section
4.12.