4
Configuring The Meter
22
4.3 USING READING CONFIGURATION (“RD.CF”)
To select the calibration mode and units of measurement, follow these steps:
1.
Press the
MENU
button until the meter shows “
RD.CF
”.
2.
Press the
T/mV
button. The meter shows one of the following:
•
“
R.1=2
”
For the two point buffer solution calibration (either 4 and 7, or 7 and 10
buffer solutions are used).
•
“
R.1=3
”
For three point buffer solution calibration (4, 7, or 7 and 10 buffer
solutions are used). (Default)
3.
Press the
/pH
button to scroll through available choices.
4.
Press the
T/mV
button to advance to the next submenu. The meter shows one
of the following:
•
“
R.2=C
”
The displayed temperature in degrees Celsius (Default)
•
“
R.2=F
”
The displayed temperature in degrees Fahrenheit.
5.
Press the
/pH
button to scroll through available choices.
6.
Press the
MENU
button to store your choice. The meter momentarily shows
“
STRD
”, followed by “
CAL.2
” or “
CAL.3
”.
Refer to step 4.1, during this step a temperature compensation type was chosen.
If “
OFF
” was chosen then proceed to the “
CAL.2
” or “
CAL.3
” menu.
If “
METR
” was selected, connect the RTD sensor’s leads to the RTD input
connector, and proceed to the “
CAL.2
” or “
CAL.3
” menu.
If “
MANU
” was selected, press the
MENU
button until the “
TEMP
” menu
displayed. (step 4.11 - Configuring Temperature).