manualshive.com logo in svg
background image

OM-USB-2404-60 User's Guide 

Functional Details 

USB connector 

The USB connector pr5 V power and communication. The voltage supplied through the USB connector 
is system-dependent, and may be less than 5 V. No external power supply is required. 

LED 

The LED indicates the device status. When connected to a USB port, the LED blinks steadily to indicate that 
the device is initialized and receiving power. Refer to the following table for the possible LED states. 

LED states 

LED State 

device status 

Off 

The device is not connected to a USB port or hub. 

Continuous single-blink 

The device is operating normally. 

Continuous double-blink 

The device is connected to a USB 1.1 Full-Speed port or hub, which may affect 
performance. 
Optimum performance requires connections to a USB 2.0 Hi-Speed host controller 
(480 Mbps) and USB 2.0 high-speed hubs. 

Strain relief slot 

Use the strain relief slot to keep the USB cable from disconnecting from the device inadvertently. Feed a tie 
wrap through the slot and secure to the USB cable when it is connected to the device.  

High-voltage applications 

For high voltage applications, we recommend using the OM-ACC-160 backshell to ensure that the terminals are 
not accessible. The backshell also provides strain relief to protect the screw terminals. 

 

Figure 4. Connector backshell (OM-ACC-160 accessory) 

Signal source connections 

You can connect ground-referenced or floating signal sources to the OM-USB-2404-60. 

 

Connect the positive signal of the signal source to the CH+ screw terminal. 

 

Connect the negative signal of the signal source to the CH – screw terminal.  

The following figures show the connections between the OM-USB-2404-60 and both a grounded signal source 
and a floating signal source. 

 

Figure 5. Connecting a grounded signal source 

Содержание OM-USB-2404-60

Страница 1: ......

Страница 2: ......

Страница 3: ...cessories 6 Unpacking 6 Installing the software 6 Installing the hardware 6 Calibrating the hardware 7 Chapter 3 Functional Details 8 Components 8 Screw terminals CH0 to CH3 8 USB connector 9 LED 9 St...

Страница 4: ...able on our website at www omega com You can also contact Omega Engineering by phone fax or email with specific questions Phone 203 359 1660 Fax 203 359 7700 Email das omega com Safety guidelines You...

Страница 5: ...sion A 250 Vrms channel to channel and channel to ground isolation protects the device and computer from ground spikes and ensures a reliable data stream An on board LED indicates the status of the de...

Страница 6: ...touching the computer chassis or other grounded object to eliminate any stored static charge If any components are missing or damaged notify Omega Engineering immediately by phone fax or e mail Phone...

Страница 7: ...e device This warm up time is required to achieve the specified rated accuracy of measurements Calibrating the hardware The OM USB 2404 60 is shipped fully calibrated Calibration coefficients are stor...

Страница 8: ...for connection to four isolated analog inputs The maximum sampling rate is 50 kS s per channel and the input voltage range is 60 V nominal The channels may be sampled individually or simultaneously Si...

Страница 9: ...peed host controller 480 Mbps and USB 2 0 high speed hubs Strain relief slot Use the strain relief slot to keep the USB cable from disconnecting from the device inadvertently Feed a tie wrap through t...

Страница 10: ...for one analog input channel Figure 7 Analog input circuitry for one channel Filtering The OM USB 2404 60 uses a combination of analog and digital filtering to provide an accurate representation of i...

Страница 11: ...artifact of signals at a higher frequency The alias free bandwidth 0 453 is defined by the ability of the filter to reject frequencies above the stopband frequency The alias free bandwidth equals the...

Страница 12: ...mum Input voltage ranges CH to CH 60 V nominal 62 64 V typical 61 5 V minimum Overvoltage protection 100 V Input coupling DC Input impedance CH to CH 1 M Input noise 320 V rms Gain drift 5 ppm C Offse...

Страница 13: ...er Table 3 Power specifications Parameter Specification Current consumption from USB 500 mA maximum Bus interface Table 4 Bus specifications Parameter Specification USB specification USB 2 0 Hi Speed...

Страница 14: ...outlet for example 115 V for US or 230 V for Europe Caution Do not connect the device to signals or use for measurements within Measurement Categories III or IV Screw terminal connectors Table 8 Screw...