2-13
2.11 Battery Pack and Power Adapter
The data logger is a portable device and can operate on battery power for several
hours before requiring recharging. The included battery is a 7.2V, 4.8 Ah Li-Ion
battery. Also included is an AC adapter to provide 12VDC power for battery
recharging.
2.11.1 Connecting the battery pack
The data logger ships with the battery installed, but disconnected to keep it from
completely draining during shipment. To connect the battery pack:
1. Remove the four screws securing the battery door using the included screw
driver.
2. Remove the battery from the compartment.
3. Plug the keyed connector on the end of the battery leads into the power
connector in the battery compartment. IMPORTANT! Observe the correct
polarity when plugging the battery into the data logger.
4. Reinstall the battery in the compartment placing the leads under the battery.
Be sure not to crimp the battery leads.
5. Reinstall the battery compartment door and tighten the 4 battery door
screws.
Hardware
2
Figure 2-26 Connecting Battery
,16(57%$77(5<
3/8*+(5(
Figure 2-25 Removing Battery Door.
,16(57%$77(5<
3/8*+(5(
Close-Up of
Battery Plug
Содержание OM-DAQXL
Страница 75: ...4 1 NOTES 4...