2-12
2.10 LED Indication
The three status LEDs on the front of the data logger provide an easy to see
visual indication of the status of the device. This is useful when the LCD has
been turned off or is difficult to read due to glare or distance. See figure 2-24 and
table 2-7 for information on the different LED indications.
Hardware
2
Figure 2-24 Status LED Locations
Led
Status
Color and State
1 – Power/
Charging
DC Adapter
Battery
Connected
Charged
Green
Connected
Charging
Green Flashing
Not connected Discharging
Green, amber battery <40% remaining,
red battery <15% remaining.
Connected
No battery
Flashing amber
2 – Logging
Logging
Flashing Green
Not logging - Error
Red
Armed mode
Amber
Free running
Green
3 - Alarm
Alarm condition
Red
No alarm
Green
Alarms disabled
Off
Table 2-7 Status LED States