![Omega OM-DAQXL Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/omega/om-daqxl/om-daqxl_user-manual_737410017.webp)
2-8
Hardware
2
2.6.2.2 Three-Wire RTD Wiring
For 3-wire RTD sensors the second common lead must be connected to the
negative terminal of the next consecutive even channel.
2.6.2.3 Four-Wire RTD Wiring
For 4-wire RTD sensors the data logger performs a Kelvin resistance
measurement using 1 channel to supply excitation voltage and the next channel
to sense the voltage across the RTD.
Figure 2-14 Three-Wire RTD Wiring Example
Figure 2-15 Four-Wire RTD Wiring Example