                                                                                                         Temperature/Event Adapter

       3

          Recording Time

              Sampling Rate

40 Days

15 Minutes

  2 Weeks

  5 Minutes

60 Hours

  1 Minute

30 Hours

30 Seconds

15 Hours

15 Seconds

              On Event Change

On to Off or, Off to On

Note that recording on event change is not a time related recording func-
tion.

 The Service Logger will monitor the Event Detector and only make a data

recording when the status of the event detector changes from ON to OFF or
from OFF to ON. 

There is also a maximum time limit between events.

 If the

Service Logger does not detect a change in state for more than 45 hours the
Service Logger will assume no events are occurring and the recording will be
automatically terminated. As in the time related recording functions the display
will continuously show the status of the temperature sensor and event detector
but will only record data on a change in the state of the event detector.

When the 

                   

M E N U

               

 

E N T E R

 button is pressed 3 more times (to cycle through the

remaining functions), the recording time and the sampling rate will be saved
into memory and the Service Logger will return to the operating mode.
Otherwise the unit will return to the operating mode after approximately 1
minute and changes will be saved into memory.

Changing Scale to °F or °C

Press the 

                   

M E N U

               

 

E N T E R

 button until the display

reads:

   

or

Use the 

              

S T A T U S

 or 

             

R E C O R D

 buttons to

change the setting to the desired scale.

When the 

                   

M E N U

               

 

E N T E R

 button is pressed twice more (to cycle through the

remaining functions), the temperature scale setting will be saved into memory,
and the Service Logger will return to the operating mode. Otherwise the unit will
return to the operating mode after approximately 1 minute and changes will be
saved into memory.

TEMPERATURE SET

TO FAHRENHEIT

TEMPERATURE SET

TO CELSIUS

Содержание OM-60-MOD-TE

Страница 1: ...D TE cables by the user is not recommended and will void the sensor warranty The voltage detector is connected by two coiled clip leads available at the interface box The voltage detector will operate...

Страница 2: ...s not connected or has become defective the display will indicate that the sensor is not responding Temperature sensor is not responding CHANGING SETTINGS Every function of the Service Logger can be s...

Страница 3: ...r and event detector but will only record data on a change in the state of the event detector When the M ENU EN TER button is pressed 3 more times to cycle through the remaining functions the recordin...

Страница 4: ...ding prior to starting a new recording The first step in making a recording is to select the proper recording time and sample rate If a sampling rate of 15 seconds is being used then every 15 seconds...

Страница 5: ...minate recording if the Service Logger was set to record previously 3 Record information to memory and to a printer This is a combination of options 1 and 2 Information is stored in memory and at the...

Страница 6: ...the upper right hand corner of the display The R indicates recording is in progress While you are recording it is possible to check how much memory has been used Press and hold the STA TU S button unt...

Страница 7: ...s is that data samples that are less than the selected temp erature change or if the state of the event detector stays the same no samples will be printed until the conditions that are programmed in t...

Страница 8: ...he temperature continued to decrease with the fan OFF at 5 32 a m on the Tuesday of the same week End of Data This indicates the end of the recording and shows that the temperature remained within 2 o...

Страница 9: ...to operate an error message will be displayed to alert the user to a printer problem There are several possible problems that may occur if the printer is not ready The Service Logger will alert the u...

Страница 10: ...have to manually scan a potential 4000 data samples to find the information of interest Press and hold M ENU EN TER until this display appears The actual recording rate is unimportant Press M ENU EN...

Страница 11: ...g of the data recording or R EC O R D to go to the last previous data sample Press M ENU EN TER at any time to return to the Operating Mode TRANSFERRING DATA TO A COMPUTER After a recording has been m...

Страница 12: ...puter screen A README file is also included on the floppy diskette which has detailed instructions on the use of the transfer program and reflects the latest information that is not included in this m...

Отзывы: