58
PART 7
APPROVALS INFORMATION
7.1 CE APPROVAL
This product conforms to the EMC directive 89/336/EEC amended by
93/68/EEC, and with the European Low Voltage Directive 72/23/EEC.
Electrical Safety EN61010-1:2001
Safety requirements for electrical equipment for measurement, control and laboratory.
Basic Insulation
Pollution Degree 2
Dielectric withstand Test per 1 min
•
Input Power to Ethernet Output:
1500Vac
•
Input Power to I/O lines (DB9):
none
Measurement Category I
Category I are measurements performed on circuits not directly connected to the Mains
Supply (power).
Transients Overvoltage Surge (1.2/50uS Pulse)
•
Input Power:
500V Transients Overvoltage
•
Ethernet:
1500V Transients Overvoltage
Note:
The ac/dc power adapter must have Safety Qualified Agency Approvals
for CE with Double Insulation rating.
The ac/dc power adapter is 9Vdc.
The minimum output current rating is 500mA.
EMC EN61000-6-1:2001 (Immunity) and EN61000-6-3:2001 (Emissions)
Immunity requirements for residential, commercial and light-industrial environments
•
EMC Emissions Table 1, Class B
•
EMC Immunity
Table 1: Enclosure
Table 2: Signal Lines Ports
Table 3: Dc input/Dc output Ports
EMC EN61326:1997 + and A1:1998 + A2:2001
Immunity and Emissions requirements for electrical equipment for measurement, control
and laboratory.
•
EMC Emissions
Table 4, Class B of EN61326
•
EMC Immunity
Table 1 of EN61326
Note:
I/O lines (DB9) require shielded cables and these cables must be located on
conductive cable trays or in conduits. Furthermore, the length of these cables
should not exceed 30 meters
Refer to the EMC and Safety installation considerations (Guidelines) of this manual
for additional information.
7.2
FCC
This device complies with Part 15, Subpart B, Class B of the FCC rules.