45 |
M 5 7 9 1
10
iServer 2 Specifications
INTERFACES
Available input ports vary depending on the iServer 2 model
Ethernet (RJ45):
1x port (Power over Ethernet available on qualifying models)
Supported Protocols:
TCP, UDP, SNMP, SNTP, ARP, ICMP, DNS, HTTP, and Telnet
Omega Link Smart Probe:
1x M12 8-Pin port
Thermocouple:
2x ports (available on qualifying models)
Digital I/O and Relays:
2x RS232/RS485 1.5 A @ 30 V DC
LED Indicators:
100 BASE-T, Network Link and Activity, DHCP, Internet
Alarm I/O:
Two contact inputs TTL 0.5; one open collector output 150 mA @ 30 V DC
Sample Rate:
8 samples/second max
Memory Capacity and Sample Rate:
The table below lists the lifespan of the internal storage before
the 11 GB of storage is filled. Some Smart Probes come with more than one active sensor. These
sensors can be individually enabled and disabled using Omega's SYNC configuration software.
Sample Rate
2 Active
Sensors
4 Active
Sensors
1 second
(max)
4 years
2 years
5 seconds
24 years
12 years
10 seconds
40 years
20 years
Management:
Device and probe configuration and monitoring through embedded WEB server
Embedded WEB Server:
Servers WEB pages containing real-time data and live updated gauge
views and charts within definable time intervals.
MECHANICAL
Dimensions of Base Device:
101.6 mm L x 155.6 mm W x 330 mm H (4 in. L x 6.13 in. W x 12.99 in. H)
not including bracket and M12 connector
Material:
Stainless Steel
Display:
LCD 32 mm L x 93.5 mm W
Weight:
655 g (1.44 lbs.), including battery
POWER
Power Input:
9 to 12 V DC
Consumption:
4 W
AC Power Adapter (Included) Nominal Output:
12 V DC @ 1.5 A
Power Over Ethernet:
IEEE 802.3AF, 44 V – 49 V, Power Consumption under 10 W
Input:
100 to 240 V AC, 50/60 Hz
Back-Up Battery:
9 V DC, alkaline. 96 hours at 5 seconds recording intervals and 1 second reading
with two connected probes
ENVIRONMENTAL
Operating Temperatures
iS
erver 2 Unit:
0 to 60°C (32 to 140°F)
Battery: